To what extent are objectively measured mammographic imaging techniques associated with compression outcomes
The British Journal of Radiology · 01.06.2023 · 96 (1146)
DOI: 10.1259/bjr.20230089 ↗Kurz gefasst
Fachartikel in der Zeitschrift The British Journal of Radiology, erschienen am 01.06.2023. Verfasst von Sue M Hudson und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Non-Communicable Disease Epidemiology, London School of Hygiene and Tropical Medicine, Keppel, London, UK. Hauptthemen laut MEDLINE: Mammography, Breast Neoplasms. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Objective
To describe the association between objectively measurable imaging techniques and the resulting compression thickness and dose.
Methods
The study included 80,495 routine screens from the South-West London Breast Screening Service between March 2013 and July 2017. Average compression force, paddle tilt and dose were calculated. The Volpara ® DensityTM algorithm was used to estimate pressure, breast volume and density. Linear regression models, using generalized estimating equations (GEEs) to account for clustering by practitioner, assessed the strength of the associations between the imaging compression outcomes, (thickness, dose) and imaging techniques (force, pressure and paddle tilt), adjusting for the subject’s characteristics (age, ethnicity, breast volume and percent mammographic density).
Results
Fully adjusted linear regression models showed that compression thickness decreased by ~1 mm (~2% of mean thickness) for every 1daN increase in force and decreased by ~0.8 mm with an increase of 1 kPa of pressure (at median pressure). Increasing pressure above 15 kPa resulted in minimal reduction in thickness. Dose increased with increased force but decreased by ~1% of mean dose with every increase in 1 kPa of pressure. For 1 o increase in paddle tilt, the compression thickness increased by ~1.5 mm (~2.5%) and dose increased by ~2.5%, (Pt
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Non-Communicable Disease Epidemiology, London School of Hygiene and Tropical Medicine, Keppel, London, UK.
- Oxford Breast Imaging Centre, Oxford University Hospitals NHS Foundation Trust, Oxford, UK.
ORCID-Profile: Sue M Hudson
Zitieren
APA
Hudson, S. M., Wilkinson, L. S., De Stavola, B. L., & Dos-Santos-Silva, I. (2023). To what extent are objectively measured mammographic imaging techniques associated with compression outcomes. The British Journal of Radiology, 96(1146). https://doi.org/10.1259/bjr.20230089
Vancouver
Hudson SM, Wilkinson LS, De Stavola BL, Dos-Santos-Silva I. To what extent are objectively measured mammographic imaging techniques associated with compression outcomes. The British Journal of Radiology. 2023;96(1146). doi:10.1259/bjr.20230089
BibTeX
@article{Hudson2023,
author = {Hudson, Sue M and Wilkinson, Louise S and De Stavola, Bianca L and Dos-Santos-Silva, Isabel},
title = {To what extent are objectively measured mammographic imaging techniques associated with compression outcomes},