Pre-TAVI aortic annulus sizing: comparison between manual and semi-automated new generation software measurements in operators with different experience
The British Journal of Radiology · 01.08.2023 · 96 (1149)
DOI: 10.1259/bjr.20220733 ↗Kurz gefasst
Fachartikel in der Zeitschrift The British Journal of Radiology, online erschienen am 25.07.2023. Verfasst von Andrea Daniele Annoni und 14 weiteren Autor:innen. Beteiligte Einrichtung: Centro Cardiologico Monzino, IRCCS, Milan, Italy. Hauptthemen laut MEDLINE: Transcatheter Aortic Valve Replacement, Aortic Valve Stenosis. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Objectives
Aim of the study is to compare manual and semi-automatic measurements for aortic annulus assessment among different operators.
Methods
Eighty patients who underwent TAVI were retrospectively enrolled. The measurements manually performed by an experienced reader for aortic annulus (minimum and maximum diameters, perimeter, area), annulus-to-coronary ostia distance and time needed for the whole evaluation, were collected. The same operator (observer 1) and two less experienced readers (observer 2 and 3, with >5 years and 1 year of experience, respectively) assessed the same measurements using a semi-automatic software. Differences between manual and semi-automatic measurements, reading time and suggested valves size derived by CT were compared.
Results
Very good correlations were found between manual and software-aided measurements for aortic annulus area and perimeter in comparison with standard measurements for the three readers (ICC range 0.81–0.98). Good correlations were found for the distance with coronary ostia(0.75–0.79). The same area-derived prosthesis size for manual and semi-automatic measurements was selected in 96% of cases for observer 1; very good correlations were also found for observer 2 and 3 (ICC = 0.89 and 0.88, respectively). Using semi-automatic measurements, the mean time needed for CT images was significantly lower for observers 1 and 2 (1.50 and 1.72versus 3.14 min), respectively.
Conclusions
Pre-TAVI CT using semi-automatic software allows accurate and reproducible measurements, reducing reconstruction time up to 50% and is reliable even for operators with different experience. Advances in knowledge: The use of semi-automatic dedicated software for CT in TAVI planning is reliable even for operators without long time experience and allows accurate and reproducible measurements improving pre-TAVI workflow.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Centro Cardiologico Monzino, IRCCS, Milan, Italy.
- Centro Cardiologico Monzino, IRCCS, University of Milan, Milan, Italy.
- ASST Grande Ospedale Metropolitano Niguarda, Milan, Italy.
- Cardiovascular Center Aalst, OLV-Clinic, Aalst, Belgium.
- Department of Clinical Sciences and Community Health, University of Milan, Milan, Italy.
Zitieren
APA
Annoni, A. D., Mancini, M. E., Carlicchi, E., Belmonte, M., Formenti, A., Mushtaq, S., Marchetti, F., Cilia, F., Baggiano, A., Fusini, L., Bonomi, A., Gennari, M., Montorsi, P., Pepi, M., & Pontone, G. (2023). Pre-TAVI aortic annulus sizing: comparison between manual and semi-automated new generation software measurements in operators with different experience. The British Journal of Radiology, 96(1149). https://doi.org/10.1259/bjr.20220733
Vancouver
Annoni AD, Mancini ME, Carlicchi E, Belmonte M, Formenti A, Mushtaq S, et al. Pre-TAVI aortic annulus sizing: comparison between manual and semi-automated new generation software measurements in operators with different experience. The British Journal of Radiology. 2023;96(1149). doi:10.1259/bjr.20220733
BibTeX
@article{Annoni2023,
author = {Annoni, Andrea Daniele and Mancini, Maria Elisabetta and Carlicchi, Eleonora and Belmonte, Marta and Formenti, Alberto and Mushtaq, Saima and Marchetti, Francesca and Cilia, Francesco and Baggiano, Andrea and Fusini, Laura and Bonomi, Alice and Gennari, Marco and Montorsi, Piero and Pepi, Mauro and Pontone, Gianluca},
title = {Pre-TAVI aortic annulus sizing: comparison between manual and semi-automated new generation software measurements in operators with different experience},
journal = {The British Journal of Radiology},
year = {2023},
volume = {96},
number = {1149},
doi = {10.1259/bjr.20220733}
}
EUVIMED