Image-mode performance characterisation of a positron emission tomography subsystem designed for Biology-guided radiotherapy (BgRT)
The British Journal of Radiology · 01.01.2023 · 96 (1141)
DOI: 10.1259/bjr.20220387 ↗Kurz gefasst
Fachartikel in der Zeitschrift The British Journal of Radiology, online erschienen am 15.11.2022. Verfasst von Zhiqiang Hu und 15 weiteren Autor:innen. Beteiligte Einrichtung: RefleXion Medical, Inc, Hayward, CA. Hauptthemen laut MEDLINE: Biology, Positron-Emission Tomography. Die Arbeit wurde bisher 18-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Objectives
In this study, we characterise the imaging-mode performance of the positron emission tomography (PET) subsystem of the RefleXion X1 machine using the NEMA NU-2 2018 standard.
Methods
The X1 machine consists of two symmetrically opposing 900 arcs of PET detectors incorporated into the architecture of a ring-gantry linear accelerator rotating up to 60 RPM. PET emissions from a tumour are detected by the PET detectors and used to guide the delivery of radiation beam. Imaging performance of the PET subsystem on X1 machine was evaluated based on sensitivity of the PET detectors, spatial resolution, count-loss performance, image quality, and daily system performance check.
Results
PET subsystem sensitivity was measured as 0.183 and 0.161 cps/kBq at the center and off-center positions, respectively. Spatial resolution: average FWHM values of 4.3, 5.1, and 6.7 mm for the point sources at 1, 10, and 20 cm off center, respectively were recorded. For count loss, max NECR: 2.63 kcps, max true coincidence rate: 5.56 kcps, and scatter fraction: 39.8%. The 10 mm sphere was not visible. Image-quality contrast values were: 29.6%, 64.9%, 66.5%, 81.8%, 81.2%, and background variability: 14.8%, 12.4%, 10.3%, 8.8%, 8.3%, for the 13, 17, 22, 28, 37 mm sphere sizes, respectively.
Conclusions
When operating in an imaging mode, the spatial resolution and image contrast of the X1 PET subsystem were comparable to those of typical diagnostic imaging systems for large spheres, while the sensitivity and count rate were lower due to the significantly smaller PET detector area in the X1 system. Clinical efficacy when used in BgRT remains to be validated. Advances in knowledge: This is the first performance evaluation of the PET subsystem on the novel BgRT machine. The dual arcs rotating PET subsystem on RefleXion X1 machine performance is comparable to those of the typical diagnostic PET system based on the spatial resolution and image contrast for larger spheres.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- RefleXion Medical, Inc, Hayward, CA.
- Department of Radiology, Stanford University, Stanford, CA.
- Department of Radiation Oncology, Stanford University, Stanford, CA.
ORCID-Profile: Murat Surucu
Zitieren
APA
Hu, Z., Bieniosek, M., Ferri, V., Iagaru, A., Kovalchuk, N., Han, B., Xing, L., Vitzthum, L., Olcott, P., Narayanan, M., Laurence, T., Ren, Y., Oderinde, O. M., Shirvani, S. M., Chang, D., & Surucu, M. (2023). Image-mode performance characterisation of a positron emission tomography subsystem designed for Biology-guided radiotherapy (BgRT). The British Journal of Radiology, 96(1141). https://doi.org/10.1259/bjr.20220387
Vancouver
Hu Z, Bieniosek M, Ferri V, Iagaru A, Kovalchuk N, Han B, et al. Image-mode performance characterisation of a positron emission tomography subsystem designed for Biology-guided radiotherapy (BgRT). The British Journal of Radiology. 2023;96(1141). doi:10.1259/bjr.20220387
BibTeX
@article{Hu2023,
author = {Hu, Zhiqiang and Bieniosek, Matthew and Ferri, Valentina and Iagaru, Andrei and Kovalchuk, Nataliya and Han, Bin and Xing, Lei and Vitzthum, Lucas and Olcott, Peter and Narayanan, Manoj and Laurence, Thomas and Ren, Yulan and Oderinde, Oluwaseyi M and Shirvani, Shervin M and Chang, Daniel and Surucu, Murat},
title = {Image-mode performance characterisation of a positron emission tomography subsystem designed for Biology-guided radiotherapy (BgRT)},
journal = {The British Journal of Radiology},
year = {2023},
volume = {96},
number = {1141},
doi = {10.1259/bjr.20220387}
}
EUVIMED