Continuous and Noninvasive Total Blood Hemoglobin Measurement Using Near-Infrared Reflectance Spectrometry

Stefan F van Wonderen, Andres Aguirre, Paul Benni, Rokus E C van den Dool, Juan Pedro Macias Pingarron, Gumersindo Javier Solares, Feras Hatib, Denise P Veelo, Alexander P J Vlaar, Zhongping Jian

Anesthesia & Analgesia · 2025

DOI: 10.1213/ane.0000000000007618 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Anesthesia & Analgesia, 2025. Verfasst von Stefan F van Wonderen und 9 weiteren Autor:innen. Beteiligte Einrichtung: From the Department of Intensive Care, Amsterdam UMC Location University of Amsterdam, Amsterdam, the Netherlands. Hauptthemen laut MEDLINE: Hemoglobins, Monitoring, Intraoperative. Die Arbeit nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Continuous measurement of total hemoglobin (tHb) can be of great clinical value during surgery, visualizing tHb trends and guiding blood transfusion practices. Devices such as those pulse oximetry-based have been available and studied over the past decade with varying results on their accuracy. We recently developed a first-of-its-kind near-infrared reflectance spectroscopy (NIRS) based algorithm to measure tHb continuously and noninvasively. This study was conducted to evaluate the performance of the NIRS based tHb algorithm using a cerebral oximetry sensor. METHODS: We performed a post hoc retrospective analysis of data collected during cardiac surgery from 3 clinical sites. Both NIRS data and blood gas tHb data were collected. Then the NIRS data was postprocessed through the NIRS based tHb algorithm to generate continuous measurement of tHb. Comparison between the 2 tHb measurements was assessed using the Bland-Altman analysis, mean absolute error (MAE), root mean square error (RMSE), 4-quadrant concordance, and error-grid analysis. RESULTS: One hundred and eighty-nine (189) patients were included in the analysis. The bias (or mean difference) and precision (or 1 standard deviation of the difference) (95% confidence interval) are 0.08 (0.02–0.19) g/dL and 1.01 (0.93–1.09) g/dL, respectively. The limits of agreement were −1.90 and 2.06 g/dL. The MAE is 0.79 (0.69–0.91) g/dL, the RMSE is 1.12 (0.94–1.30 g/dL), and the concordance is 86.1 (79.8–92.1) %. The error-grid analysis demonstrated that the majority (84%) of the measurements are in the green zone and 0% in the red zone. CONCLUSIONS: The exploratory study shows that the NIRS based tHb provided an accurate measurement of tHb in cardiac surgery patients. Further research may be needed to evaluate its accuracy and implementation in other clinical settings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HemoglobinsMonitoring, IntraoperativeHumansSpectroscopy, Near-InfraredRetrospective StudiesMaleFemaleOximetryMiddle AgedAgedCardiac Surgical ProceduresAlgorithmsReproducibility of ResultsBiomarkersPredictive Value of Tests

Substanzen

HemoglobinsBiomarkers

Autor:innen und Institutionen

  • From the Department of Intensive Care, Amsterdam UMC Location University of Amsterdam, Amsterdam, the Netherlands.
  • Advanced Patient Monitoring, Becton, Dickinson and Company, Irvine, California.
  • Department of Anesthesiology, Amsterdam UMC Location University of Amsterdam, Amsterdam, the Netherlands.
  • Department of Anesthesiology, Badajoz University Hospital, Badajoz, Spain.
  • Department of Anesthesiology, Hospital Universitario Marqués de Valdecilla, Santander, Spain.

Zitieren

APA

van Wonderen, S. F., Aguirre, A., Benni, P., van den Dool, R. E. C., Macias Pingarron, J. P., Solares, G. J., Hatib, F., Veelo, D. P., Vlaar, A. P. J., & Jian, Z. (2025). Continuous and Noninvasive Total Blood Hemoglobin Measurement Using Near-Infrared Reflectance Spectrometry. Anesthesia & Analgesia. https://doi.org/10.1213/ane.0000000000007618

Vancouver

van Wonderen SF, Aguirre A, Benni P, van den Dool REC, Macias Pingarron JP, Solares GJ, et al. Continuous and Noninvasive Total Blood Hemoglobin Measurement Using Near-Infrared Reflectance Spectrometry. Anesthesia & Analgesia. 2025. doi:10.1213/ane.0000000000007618

BibTeX
@article{vanWonderen2025,
  author = {van Wonderen, Stefan F and Aguirre, Andres and Benni, Paul and van den Dool, Rokus E C and Macias Pingarron, Juan Pedro and Solares, Gumersindo Javier and Hatib, Feras and Veelo, Denise P and Vlaar, Alexander P J and Jian, Zhongping},
  title = {Continuous and Noninvasive Total Blood Hemoglobin Measurement Using Near-Infrared Reflectance Spectrometry},
  journal = {Anesthesia & Analgesia},
  year = {2025},
  doi = {10.1213/ane.0000000000007618}
}
Kostenlos starten →