Ex vivo mechanical testing of double strand, braided and knitted polyethylene suture for acute transverse section of the Achilles tendon in a dog model

Manar Y Abd El-Aziz, Doaa H Elgohary, Y A Abo El Amaim, Elham A Hassan

Journal of Orthopaedic Surgery and Research · 2025

DOI: 10.1186/s13018-025-05947-1 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Orthopaedic Surgery and Research, 2025. Verfasst von Manar Y Abd El-Aziz und 3 weiteren Autor:innen. Beteiligte Einrichtung: Clothing and Knitting Industrial Research Department, Textile Research and Technology Institute, National Research Centre, 33 EL Bohouth St. (former …. Hauptthemen laut MEDLINE: Achilles Tendon, Sutures, Polyethylene, Suture Techniques. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Cairo University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Acute tendon cut represents a great challenge both in human and veterinary medical practice. The current study aimed to compare the ultimate biomechanical properties (tensile strength, elongation, stress load, yield load and break load) of double strand, braided and knitted polyethylene suture in an ex vivo model of acute transverse section of the Achilles tendon in dog model using locking loop suture and three-loop pulley suture.

Methods

A-thirty-six Achilles tendon was transected from 18 dog cadavers. Tendon samples were randomly allocated (6 tendons/group) to be sutured either by double strand, braided and novel knitted formation techniques from polyethylene suture using either three-loop pulley suture or locking loop suture patterns. Biomechanical testing of different yarn for tensile strength, elongation, stress, yield load, break load was performed.

Results

Braided polyethylene sutures demonstrated superior biomechanical properties, showing the highest maximum tension, load, stress, and yield load, while knitted sutures exhibited the greatest strain and elongation due to their looped structure. Despite the knitted yarn’s high elongation, its tensile strength and load-bearing capabilities were significantly lower. Overall, yarn formation had a greater influence on biomechanical performance in association with suturing technique. The three-loop pulley suturing demonstrated significantly improved suturing outcomes.

Conclusion

Both novel knitted, and braided suture structure demonstrated improved biomechanical properties of tendon suturing by increasing the number of strands within the tendon, simplifying the suturing process, reducing the needle passes, and minimizing tendon punctures that may interfere with healing and the overall strength. Suturing technique had a major influence on the biomechanical properties where the three-loop pulley suture demonstrated superior biomechanical properties compared to locking loop suturing.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Achilles TendonSuturesPolyethyleneSuture TechniquesMaterials TestingMechanical TestsAnimalsDogsTensile StrengthBiomechanical PhenomenaCadaverModels, Animal

Stichwörter der Autor:innen

Animal modelPolyethylene yarnSutureTendon repairTenorrhaphy

Substanzen

Polyethylene

Autor:innen und Institutionen

  • Clothing and Knitting Industrial Research Department, Textile Research and Technology Institute, National Research Centre, 33 EL Bohouth St. (former EL Tahrir St.)- Dokki, Giza, 12622, Egypt.
  • Spinning and Weaving Engineering Department, Textile Research and Technology Institute, National Research Centre, 33 EL Bohouth St. (former EL Tahrir St.)- Dokki, Giza, 12622, Egypt.
  • Spinning, Weaving and Knitting Department, Faculty of Applied Arts, Beni Suef University, Beni Suef, 62512, Egypt.
  • Department of Surgery, Anesthesiology and Radiology- Faculty of Veterinary Medicine, Cairo University, Giza, 12211, Egypt. elhamhassan@cu.edu.eg.

Förderung

  • Cairo University

Zitieren

APA

Abd El-Aziz, M. Y., Elgohary, D. H., Abo El Amaim, Y. A., & Hassan, E. A. (2025). Ex vivo mechanical testing of double strand, braided and knitted polyethylene suture for acute transverse section of the Achilles tendon in a dog model. Journal of Orthopaedic Surgery and Research. https://doi.org/10.1186/s13018-025-05947-1

Vancouver

Abd El-Aziz MY, Elgohary DH, Abo El Amaim YA, Hassan EA. Ex vivo mechanical testing of double strand, braided and knitted polyethylene suture for acute transverse section of the Achilles tendon in a dog model. Journal of Orthopaedic Surgery and Research. 2025. doi:10.1186/s13018-025-05947-1

BibTeX
@article{AbdElAziz2025,
  author = {Abd El-Aziz, Manar Y and Elgohary, Doaa H and Abo El Amaim, Y A and Hassan, Elham A},
  title = {Ex vivo mechanical testing of double strand, braided and knitted polyethylene suture for acute transverse section of the Achilles tendon in a dog model},
  journal = {Journal of Orthopaedic Surgery and Research},
  year = {2025},
  doi = {10.1186/s13018-025-05947-1}
}
Kostenlos starten →