Time-elapsed microstructural imaging of failure of the reverse shoulder implant
Journal of Orthopaedic Surgery and Research · 2024
DOI: 10.1186/s13018-024-04652-9 ↗Kurz gefasst
Fallbericht in der Zeitschrift Journal of Orthopaedic Surgery and Research, 2024. Verfasst von Saulo Martelli und 4 weiteren Autor:innen. Beteiligte Einrichtung: School of Mechanical Medical and Process Engineering, Queensland University of Technology, Gardens Point Campus, P'Block, Level 7, Room 717, Brisbane…. Hauptthemen laut MEDLINE: Shoulder Joint, Fractures, Bone. Die Arbeit nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Australian Research Council.
Abstract (Originalsprache)
Background
Reverse Shoulder Arthroplasties (RSA) have become a primary choice for improving shoulder function and pain. However, the biomechanical failure mechanism of the humeral component is still unclear. The present study reports a novel protocol for microstructural imaging of the entire humerus implant under load before and after fracture.
Methods
A humerus specimen was obtained from a 75-year-old male donor. An expert surgeon implanted the specimen with a commonly used RSA implant (Aequalis reversed II, Stryker Orthopaedics, USA) and surgical procedure. The physiological glenohumeral contact force that maximized the distal implant migration was selected from a public repository ( orthoload.com ). Imaging and concomitant mechanical testing were performed using a large-volume micro-CT scanner (Nikon XT H 225 ST) and a custom-made compressive stage. Both when intact and once implanted, the specimen was tested under a pre-load and by imposing a constant deformation causing a physiological reaction load (650 N, 10 degrees adducted). The deformation of the implanted specimen was then increased up to fracture, which was identified by a sudden drop of the reaction force, and the specimen was then re-scanned.
Results
The specimen’s stiffness decreased from 874 N/mm to 464 N/mm after implantation, producing movements of the bone-implant interface consistent with the implant’s long-term stability reported in the literature. The micro-CT images displayed fracture of the tuberosity, caused by a combined compression and circumferential tension, induced by the distal migration of the implant.
Conclusion
The developed protocol offers detailed information on implant mechanics under load relative to intact conditions and fracture, providing insights into the failure mechanics of RSA implants. This protocol can be used to inform future implant design and surgical technique improvements.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Mechanical Medical and Process Engineering, Queensland University of Technology, Gardens Point Campus, P'Block, Level 7, Room 717, Brisbane, QLD, 4000, Australia. saulo.martelli@qut.edu.au.
- Medical Devices Research Institute, College of Science and Engineering, Flinders University, Adelaide, SA, Australia. saulo.martelli@qut.edu.au.
- Queensland Unit for Advanced Shoulder Research (QUASR), Queensland University of Technology, Brisbane, QLD, Australia. saulo.martelli@qut.edu.au.
- Medical Devices Research Institute, College of Science and Engineering, Flinders University, Adelaide, SA, Australia.
- School of Mechanical Medical and Process Engineering, Queensland University of Technology, Gardens Point Campus, P'Block, Level 7, Room 717, Brisbane, QLD, 4000, Australia.
- Greenslopes Private Hospital, Brisbane, QLD, Australia.
Förderung
- Australian Research Council
Zitieren
APA
Martelli, S., Perilli, E., Fan, X., Rapagna, S., & Gupta, A. (2024). Time-elapsed microstructural imaging of failure of the reverse shoulder implant. Journal of Orthopaedic Surgery and Research. https://doi.org/10.1186/s13018-024-04652-9
Vancouver
Martelli S, Perilli E, Fan X, Rapagna S, Gupta A. Time-elapsed microstructural imaging of failure of the reverse shoulder implant. Journal of Orthopaedic Surgery and Research. 2024. doi:10.1186/s13018-024-04652-9
BibTeX
@article{Martelli2024,
author = {Martelli, Saulo and Perilli, Egon and Fan, Xiaolong and Rapagna, Sophie and Gupta, Ashish},
title = {Time-elapsed microstructural imaging of failure of the reverse shoulder implant},
journal = {Journal of Orthopaedic Surgery and Research},
year = {2024},
doi = {10.1186/s13018-024-04652-9}
}
EUVIMED