EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

The Aortic Magna Ease Bioprosthesis: Exploring Hemodynamic Performance Across Sexes and Prosthesis Sizes

Eliza Calin, Andréanne Cartier, Olivier Nguyen, Roxanne St-Louis, Dimitri Kalavrouziotis, Nicolas Noiseux, Christine Pacheco, Louis-Mathieu Stevens, Siamak Mohammadi, Jessica Forcillo

Journal of the Heart Valve Society · 2025

DOI: 10.1177/30494826241297119 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of the Heart Valve Society, 2025. Verfasst von Eliza Calin und 9 weiteren Autor:innen. Die Arbeit nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Edwards Lifesciences.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

This study aims to assess the incidence of prosthesis–patient mismatch (PPM), the hemodynamic performance, as well as the clinical outcomes of the Carpentier-Edwards Magna Ease (CEME) bioprosthesis in the aortic position while addressing sex-based disparities and variations in valve size.

Methods

Between 2012 and 2022, 1062 patients underwent an AVR with a CEME bioprosthesis in two high-volume cardiac surgery centers. Predicted PPM was assessed using manufacturer and Lancellotti et al. chart values, alongside measured values. Echocardiographic measurements were evaluated over a 5-year period.

Results

Predicted PPM based on the manufacturer chart was noted in 2.1% of women and 0.28% of men ( P < .001). When employing definitions by Lancellotti et al., PPM was detected in 45% of females and 18% of males ( P < .001). Notable differences were observed in terms of severe measured PPM in favor of male patients (F: 30%, M: 17%, P < .01). Regarding prosthesis sizes, the comparison of PPM from the manufacturer's chart with Lancellotti et al. values revealed significant disparities: 2% versus 58% in the 19 to 21 mm group, 0.27% versus 13% in the 23 mm group, and 0% versus 1.1% in the ≥25 mm valve sizes ( P < .001). Measured severe PPM was encountered in 30% of patients with 19 to 21 mm prosthesis, 20% in the 23 mm group and 12% in the >25 mm prosthesis. Despite these differences, mortality rates remained comparable across sexes and valve sizes.

Conclusion

Female patients and those with 19 to 21 mm prostheses experience more severe PPM compared to men or individuals with larger valves. In this contemporary cohort, PPM does not impact short-term mortality.

Förderung

  • Edwards Lifesciences

Zitieren

APA

Calin, E., Cartier, A., Nguyen, O., St-Louis, R., Kalavrouziotis, D., Noiseux, N., Pacheco, C., Stevens, L., Mohammadi, S., & Forcillo, J. (2025). The Aortic Magna Ease Bioprosthesis: Exploring Hemodynamic Performance Across Sexes and Prosthesis Sizes. Journal of the Heart Valve Society. https://doi.org/10.1177/30494826241297119

Vancouver

Calin E, Cartier A, Nguyen O, St-Louis R, Kalavrouziotis D, Noiseux N, et al. The Aortic Magna Ease Bioprosthesis: Exploring Hemodynamic Performance Across Sexes and Prosthesis Sizes. Journal of the Heart Valve Society. 2025. doi:10.1177/30494826241297119

BibTeX
@article{Calin2025,
  author = {Calin, Eliza and Cartier, Andréanne and Nguyen, Olivier and St-Louis, Roxanne and Kalavrouziotis, Dimitri and Noiseux, Nicolas and Pacheco, Christine and Stevens, Louis-Mathieu and Mohammadi, Siamak and Forcillo, Jessica},
  title = {The Aortic Magna Ease Bioprosthesis: Exploring Hemodynamic Performance Across Sexes and Prosthesis Sizes},
  journal = {Journal of the Heart Valve Society},
  year = {2025},
  doi = {10.1177/30494826241297119}
}
Kostenlos starten →