EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

Off-label Use of the Aviator Plus Balloon in Infants with Left Heart Obstruction: A Single-center Early Experience

Anil Kumar Singhi, Anish Nath, Arnab De, Kaushik Mukherjee, Kasturi H. Bandyopadhyay, Hirak Suvra Majumdar

Indian Journal of Clinical Cardiology · 2026-09 · 7 (3) · 222-229

DOI: 10.1177/26324636261438608 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Indian Journal of Clinical Cardiology, online erschienen am 09.04.2026. Verfasst von Anil Kumar Singhi und 5 weiteren Autor:innen. Die Arbeit nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Balloon dilation is an established option for selected left heart obstructive lesions in infants, traditionally using low-profile, low-pressure balloons. Experience with high-pressure, rapid-exchange systems such as the Aviator Plus in this age group remains limited.

Objective

To describe feasibility, procedural performance, and early outcomes of off-label balloon dilation using the Aviator Plus platform in infants with coarctation of the aorta or valvular aortic stenosis.

Methods

A retrospective review was performed of infants undergoing Aviator Plus balloon dilation between April 2024 and July 2025. Clinical features, procedural details, and follow-up outcomes were summarized descriptively.

Results

Four infants (mean age 5.5 months; mean weight 5.87 kg) underwent intervention: three with native coarctation of the aorta and one with severe aortic stenosis. One coarctation of the aorta patient had significant left ventricular dysfunction; the aortic stenosis patient had moderate mitral regurgitation. Balloon dilation used a single 6 × 20 mm balloon in two cases and sequential 6 × 20 mm and 7 × 20 mm balloons in two. All procedures produced immediate gradient reduction. During a mean 8.5-month follow-up, one infant with aortic stenosis died from progression of mitral regurgitation and intercurrent infection, and one coarctation of the aorta patient required surgical repair for re-coarctation. The remaining infants maintained gradient relief with preserved ventricular function.

Conclusions

This early experience suggests that the Aviator Plus balloon can be a feasible off-label option for carefully selected infants with left heart obstruction, particularly where pediatric-specific balloons are limited. Larger series with extended follow-up are needed to define long-term outcomes.

Zitieren

APA

Singhi, A. K., Nath, A., De, A., Mukherjee, K., Bandyopadhyay, K. H., & Majumdar, H. S. (2026). Off-label Use of the Aviator Plus Balloon in Infants with Left Heart Obstruction: A Single-center Early Experience. Indian Journal of Clinical Cardiology, 7(3), 222-229. https://doi.org/10.1177/26324636261438608

Vancouver

Singhi AK, Nath A, De A, Mukherjee K, Bandyopadhyay KH, Majumdar HS. Off-label Use of the Aviator Plus Balloon in Infants with Left Heart Obstruction: A Single-center Early Experience. Indian Journal of Clinical Cardiology. 2026;7(3):222-229. doi:10.1177/26324636261438608

BibTeX
@article{Singhi2026,
  author = {Singhi, Anil Kumar and Nath, Anish and De, Arnab and Mukherjee, Kaushik and Bandyopadhyay, Kasturi H. and Majumdar, Hirak Suvra},
  title = {Off-label Use of the Aviator Plus Balloon in Infants with Left Heart Obstruction: A Single-center Early Experience},
  journal = {Indian Journal of Clinical Cardiology},
  year = {2026},
  volume = {7},
  number = {3},
  pages = {222--229},
  doi = {10.1177/26324636261438608}
}
Kostenlos starten →