Assessing Body Composition for Resource-Limited Settings: Agreement of a Scalable BIA Device Compared to DXA and Clinical BIA in Adults With Obesity
Journal of Primary Care & Community Health · 2026-06 · 17
DOI: 10.1177/21501319261459756 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Primary Care & Community Health, online erschienen am 11.06.2026. Verfasst von Katie M Ellison und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Family and Community Medicine, University of Alabama at Birmingham, Birmingham, AL, USA. Hauptthemen laut MEDLINE: Absorptiometry, Photon, Electric Impedance, Body Composition, Obesity. Die Arbeit nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: NIDDK NIH HHS, University of Alabama at Birmingham (UAB) Nutrition Obesity Research Center, National Institutes of Health, National Institute for Diabetes, and Digestive, and Kidney Diseases.
Abstract (Originalsprache)
Background
Accurate tracking of skeletal muscle mass (SMM) and fat mass percent (FM%) is central to evaluating obesity treatment. The InBody H20N bioelectrical impedance analysis (BIA) device could offer a scalable, lower-cost alternative. However, its accuracy compared to dual-energy X-ray absorptiometry (DXA) and research-grade BIA in adults with obesity is unknown.
Methods
Baseline SMM and FM% was measured in 55 participants with obesity (62.6±7.6 y, 67.3% female, BMI: 37.6±5.7 kg/m 2 ) enrolled in a behavioral weight-loss program using three modalities: the InBody H20N, InBody S10, and DXA. Agreement was assessed using intraclass correlation coefficients (ICC) analyses with two-way mixed-effects and absolute-agreement, Bland-Altman plots with limits of agreement, and Lin’s concordance correlation coefficients.
Results
For SMM, agreement was good between H20N and DXA (ICC 0.78; 95% CI -0.15-0.94) and excellent between H20N and S10 (ICC 0.93; 95% CI 0.87–0.96). For FM%, agreement was excellent between H20N and DXA (ICC 0.93; 95% CI 0.79–0.97) and good between H20N and S10 (ICC 0.81; 95% CI 0.66–0.89). All ICCs were statistically significant (p
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Family and Community Medicine, University of Alabama at Birmingham, Birmingham, AL, USA.
ORCID-Profile: Katie M Ellison
Förderung
- NIDDK NIH HHS
- University of Alabama at Birmingham (UAB) Nutrition Obesity Research Center
- National Institutes of Health, National Institute for Diabetes, and Digestive, and Kidney Diseases
Zitieren
APA
Ellison, K. M., Gard, W., Wood, J., Layton, C., & Sayer, R. D. (2026). Assessing Body Composition for Resource-Limited Settings: Agreement of a Scalable BIA Device Compared to DXA and Clinical BIA in Adults With Obesity. Journal of Primary Care & Community Health, 17. https://doi.org/10.1177/21501319261459756
Vancouver
Ellison KM, Gard W, Wood J, Layton C, Sayer RD. Assessing Body Composition for Resource-Limited Settings: Agreement of a Scalable BIA Device Compared to DXA and Clinical BIA in Adults With Obesity. Journal of Primary Care & Community Health. 2026;17. doi:10.1177/21501319261459756
BibTeX
@article{Ellison2026,
author = {Ellison, Katie M and Gard, William and Wood, Jenna and Layton, Chase and Sayer, R Drew},
title = {Assessing Body Composition for Resource-Limited Settings: Agreement of a Scalable BIA Device Compared to DXA and Clinical BIA in Adults With Obesity},
journal = {Journal of Primary Care & Community Health},
year = {2026},
volume = {17},
doi = {10.1177/21501319261459756}
}