Adaptation of the clinical limits of use tool (CLOUT): Clinical and environmental testing of off-road mobility devices: A case study

Brandon Daveler iD, Sangmi Park iD, Benjamin Gebrosky, Rosemarie Cooper, Rory Cooper, Kainat Shahid, Kendra Betz, Ian Rice iD

Journal of Rehabilitation and Assistive Technologies Engineering · 2026-08 · 13

DOI: 10.1177/20556683261478129 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Rehabilitation and Assistive Technologies Engineering, online erschienen am 12.08.2026. Verfasst von Brandon Daveler und 7 weiteren Autor:innen. Beteiligte Einrichtung: Human Engineering Research Laboratories, Department of Veterans Affairs, VA Pittsburgh Healthcare System, Pittsburgh, Pennsylvania, USA. Die Arbeit nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: National Institute on Disability, Independent Living, and Rehabilitation Research.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study presents a case-based application of the Clinical Limits of Use Tool (CLOUT), specifically its clinical device testing component, to evaluate an off-road mobility device, the Invacare Top End Force CC handcycle. This component of CLOUT was operationalized into measurable variables and standardized test procedures through collaboration among therapists, engineers, and advisory expert users. The evaluation included structured interviews and controlled off-road testing assessing assembly, seating and positioning, transfers, obstacle negotiation, slope performance, terrain interaction, and safety. Results The results demonstrated that the handcycle performed effectively on moderate terrain, including dirt, gravel, and grass, and successfully navigated slopes up to approximately 29° and smaller obstacles. Performance limitations were identified in larger obstacle negotiation due to low ground clearance, as well as in transfers due to low seat height and limited support. Device performance was influenced by user strength and environmental conditions. This case study demonstrates how CLOUT can be applied as a structured and reproducible framework to evaluate mobility device performance in non-clinical environments. The findings highlight the importance of aligning user ability, device characteristics, and environmental demands to support safe and effective mobility. This approach may inform clinical decision-making, user training, and future development of standardized evaluation methods for off-road mobility technologies and beyond.

Schlagwörter

Stichwörter der Autor:innen

Clinical Limits of Use Tool (CLOUT)device limitsmobility devicessafetyuser needs

Autor:innen und Institutionen

  • Human Engineering Research Laboratories, Department of Veterans Affairs, VA Pittsburgh Healthcare System, Pittsburgh, Pennsylvania, USA.
  • Department of Physical Medicine and Rehabilitation, University of Pittsburgh, Pittsburgh, Pennsylvania, USA.
  • Department of Rehabilitation Science and Technology, University of Pittsburgh, Pittsburgh, Pennsylvania, USA.
  • Department of Bioengineering, University of Pittsburgh, Pittsburgh, Pennsylvania, USA.
  • Department of Health and Kinesiology, University of Illinois Urbana Champaign, Champaign, Illinois.
  • VA National Center for Patient Safety, Department of Veterans Affairs, Ann Arbor, Michigan, USA.

ORCID-Profile: Brandon Daveler, Sangmi Park, Ian Rice

Förderung

  • National Institute on Disability, Independent Living, and Rehabilitation Research

Zitieren

APA

Daveler, B., Park, S., Gebrosky, B., Cooper, R., Cooper, R., Shahid, K., Betz, K., & Rice, I. (2026). Adaptation of the clinical limits of use tool (CLOUT): Clinical and environmental testing of off-road mobility devices: A case study. Journal of Rehabilitation and Assistive Technologies Engineering, 13. https://doi.org/10.1177/20556683261478129

Vancouver

Daveler B, Park S, Gebrosky B, Cooper R, Cooper R, Shahid K, et al. Adaptation of the clinical limits of use tool (CLOUT): Clinical and environmental testing of off-road mobility devices: A case study. Journal of Rehabilitation and Assistive Technologies Engineering. 2026;13. doi:10.1177/20556683261478129

BibTeX
@article{Daveler2026,
  author = {Daveler, Brandon and Park, Sangmi and Gebrosky, Benjamin and Cooper, Rosemarie and Cooper, Rory and Shahid, Kainat and Betz, Kendra and Rice, Ian},
  title = {Adaptation of the clinical limits of use tool (CLOUT): Clinical and environmental testing of off-road mobility devices: A case study},
  journal = {Journal of Rehabilitation and Assistive Technologies Engineering},
  year = {2026},
  volume = {13},
  doi = {10.1177/20556683261478129}
}
Kostenlos starten →