Fachartikel in der Zeitschrift Journal of Rehabilitation and Assistive Technologies Engineering, online erschienen am 04.12.2020. Verfasst von Matthias Panny und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechatronics, MCI, University of Applied Sciences, Innsbruck, Austria. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Introduction
Available robot-assisted stroke rehabilitation systems are often limited in their utilization in the home environment, due to several barriers such as high cost, absence of therapists, tedious training tasks, or encumbering interfaces. This paper presents a low-cost robotic rehabilitation and assessment device for restoring wrist function, offering wrist exercises incorporating pronation-supination and flexion-extension movements. Furthermore, the device is designed for the assessment of joint stiffness of the wrist, and range of motion in two degrees of freedom.
Methods
Mechanical/electrical design of the device as well as the control system is described. A preliminary evaluation focused on the measurement of the torsional stiffness of the limb is presented. It is evaluated by reconstructing the known stiffness values of torsional springs by measuring the motor current required to displace them.
Results
The device demonstrates the ability to determine the stiffness of an object with low-cost hardware. Use case scenarios of the device for training and assessment of the wrist are presented, allowing for a range of motion of ± 75 ° and ± 65 ° , for pronation-supination and flexion-extension respectively.
Conclusion
The device shows potential to help objectively quantify the stiffness of the wrist movement, which consecutively could be used to represent and quantify the degree of impairment of patients after stroke in a more objective manner. Further clinical study is necessary to examine this.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Mechatronics, MCI, University of Applied Sciences, Innsbruck, Austria.
- Department of Neurology, Hospital Hochzirl-Natters, Zirl, Austria.
ORCID-Profile: Matthias Panny
Zitieren
APA
Panny, M., Mayr, A., Nagiller, M., & Kim, Y. (2020). A domestic robotic rehabilitation device for assessment of wrist function for outpatients. Journal of Rehabilitation and Assistive Technologies Engineering, 7. https://doi.org/10.1177/2055668320961233
Vancouver
Panny M, Mayr A, Nagiller M, Kim Y. A domestic robotic rehabilitation device for assessment of wrist function for outpatients. Journal of Rehabilitation and Assistive Technologies Engineering. 2020;7. doi:10.1177/2055668320961233
BibTeX
@article{Panny2020,
author = {Panny, Matthias and Mayr, Andreas and Nagiller, Marco and Kim, Yeongmi},
title = {A domestic robotic rehabilitation device for assessment of wrist function for outpatients},
journal = {Journal of Rehabilitation and Assistive Technologies Engineering},
year = {2020},
volume = {7},
doi = {10.1177/2055668320961233}
}
EUVIMED