Utilization of lower compliance voltages for effective clinical neuromuscular electrical stimulation

K M Bogie iD, R Kobetic, D B Shire

Journal of Rehabilitation and Assistive Technologies Engineering · 2019-01 · 6

DOI: 10.1177/2055668319847116 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Rehabilitation and Assistive Technologies Engineering, online erschienen am 28.06.2019. Verfasst von K M Bogie und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedics, Case Western Reserve University, Cleveland, OH, USA. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Regular use of neuromuscular electrical stimulation (NMES) for functional applications can produce changes in stimulated muscles that increase soft tissue health. This has spurred numerous attempts to develop clinical stimulation systems for weight-shifting and pressure injury (PrI) prevention, for example in wheelchairbound individuals who use their chairs for hours at a time. For these individuals, stimulation of the gluteus muscles and subsequent contraction can cause weight shifts that can substantially reduce PrI development. Surface gluteal electrical stimulation (ES) was first reported and patented over 25 years ago. Unfortunately, surface NMES has limited efficacy in providing repeated and reliable gluteal contractile responses both clinically and practically. The inferior gluteal nerve lies relatively deep to the buttock surface and close to the sciatic nerve. Reliable surface electrode placement in the upper buttock region for preferential recruitment of the inferior gluteal nerve is difficult to achieve. Repeated application of electrodes to the same area of skin can in itself cause skin damage. In previous work, study participants commented that they were motivated to join a study to evaluate the use of surface ES because they felt there was no effective approach for pressure injury prevention currently available to them. However, they found the surface NMES system cumbersome and reported that daily placement of the electrodes was time-consuming. They stated that they would be unlikely to consider long-term use of a surface NMES system due to these limitations. While use of a custom-fitted garment may alleviate some placement issues; these systems take more than 15min to don or doff. Weight gain makes any fitted garment challenging to put on and this is exacerbated for individuals with limited mobility. Any increase in subcutaneous fat will also raise the stimulus current required to achieve a contractile response. No surface NMES systems have enabled automated weight-shifting. These interventions have failed to become standard of care and associated patents have lapsed. Implanted stimulation systems for long-term therapeutic use have dual advantages. The stimulating tip of the electrode can be located very close to the motor point of the nerve of interest. This reduces the charge required to elicit a contractile response and ensures that the response is repeatable and predictable. In addition, the user does not have to apply and remove the stimulating electrodes. Weight gain over time will not alter the responses, since the stimulating tip is adjacent to the nerve. Implanted systems avoid potential repetitive damage to fragile skin and enable use on demand or continuously throughout the day. An implanted system is reliable and simple to use on a day-to-day basis and over the longer term in order to maintain hypertrophy of paralyzed gluteal muscles. Early implanted stimulation systems included onboard battery power, which necessitated surgical replacement of the entire implanted system when the battery power started to drain. This was recognized as a limitation and mathematical models were developed to reduce the rate of power consumption. Other groups developed wireless, inductively powered stimulators which removed the limitation of loss of power and the associated need for revision surgery. When considering NMES for intramuscular (IM) stimulation, it is essential that the stimulator have sufficient power to reliably generate selective action potentials and elicit a contractile response. The output must Journal of Rehabilitation and Assistive Technologies Engineering Volume 6: 1–6 ! The Author(s) 2019 Article reuse guidelines: sagepub.com/journals-permissions DOI: 10.1177/2055668319847116 journals.sagepub.com/home/jrt

Autor:innen und Institutionen

  • Department of Orthopaedics, Case Western Reserve University, Cleveland, OH, USA.
  • Louis Stokes Cleveland Department of Veterans Affairs Medical Center (LSCVAMC), Cleveland, OH, USA.
  • Department of Electrical and Computer Engineering, Cornell University, Ithaca, NY, USA.

ORCID-Profile: K M Bogie

Zitieren

APA

Bogie, K. M., Kobetic, R., & Shire, D. B. (2019). Utilization of lower compliance voltages for effective clinical neuromuscular electrical stimulation. Journal of Rehabilitation and Assistive Technologies Engineering, 6. https://doi.org/10.1177/2055668319847116

Vancouver

Bogie KM, Kobetic R, Shire DB. Utilization of lower compliance voltages for effective clinical neuromuscular electrical stimulation. Journal of Rehabilitation and Assistive Technologies Engineering. 2019;6. doi:10.1177/2055668319847116

BibTeX
@article{Bogie2019,
  author = {Bogie, K M and Kobetic, R and Shire, D B},
  title = {Utilization of lower compliance voltages for effective clinical neuromuscular electrical stimulation},
  journal = {Journal of Rehabilitation and Assistive Technologies Engineering},
  year = {2019},
  volume = {6},
  doi = {10.1177/2055668319847116}
}
Kostenlos starten →