Monitoring activity of inpatient lower limb prosthetic users in rehabilitation using accelerometry: Validation study

Salih A Salih, Nancye M Peel, Kara Burgess

Journal of Rehabilitation and Assistive Technologies Engineering · 2016

DOI: 10.1177/2055668316642387 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Rehabilitation and Assistive Technologies Engineering, 2016. Verfasst von Salih A Salih und 2 weiteren Autor:innen. Beteiligte Einrichtung: The University of Queensland School of Medicine, Princess Alexandra Hospital, Brisbane, Australia. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Due to limited accuracy of self-reported activities of lower limb prosthetic users, there is increasing interest in providing accurate walking time for those who are mobilising using their prosthesis in the rehabilitation ward. The aim of this study was to test the accuracy of a tri axial accelerometer (ActivPAL) in measuring walking activity of amputee patients while using their prosthetic limb.

Method

For the study, 21 subjects wore accelerometer devices taped to the thigh on both the amputated and non-amputated side. Each subject was asked to perform a set of activity routines: walking with prosthesis for 5 minutes, self-propelling in a wheel chair for 3 minutes and being pushed in a wheel chair for 3 minutes. Each activity was observed and timed by a physiotherapist and the observed times were compared with the output from accelerometer monitoring.

Results

Using the Bland–Altman method, the mean difference between observed and ActivPal monitor for total time spent walking for the non-amputated side was 0.004 seconds (limits of agreement −0.09 to 0.10 seconds) and for amputated side was 0.11 seconds (limits of agreement −0.43 to 0.66 seconds). An analysis between monitored and observed time found the sensitivity for the non-amputated side was 90.5% and 86% for the amputated side.

Conclusion

The use of ActivPAL accelerometers in measuring walking time for an individual using a prosthesis is within acceptable levels of accuracy for continuous ambulation monitoring. It has potentially important clinical application for prescribing prostheses.

Schlagwörter

Stichwörter der Autor:innen

Age-related rehabilitationamputeesdisabilitygait rehabilitationlimb prostheticsrehabilitation

Autor:innen und Institutionen

  • The University of Queensland School of Medicine, Princess Alexandra Hospital, Brisbane, Australia.
  • Queensland Health, Princess Alexandra Hospital, Brisbane, Australia.

Zitieren

APA

Salih, S. A., Peel, N. M., & Burgess, K. (2016). Monitoring activity of inpatient lower limb prosthetic users in rehabilitation using accelerometry: Validation study. Journal of Rehabilitation and Assistive Technologies Engineering. https://doi.org/10.1177/2055668316642387

Vancouver

Salih SA, Peel NM, Burgess K. Monitoring activity of inpatient lower limb prosthetic users in rehabilitation using accelerometry: Validation study. Journal of Rehabilitation and Assistive Technologies Engineering. 2016. doi:10.1177/2055668316642387

BibTeX
@article{Salih2016,
  author = {Salih, Salih A and Peel, Nancye M and Burgess, Kara},
  title = {Monitoring activity of inpatient lower limb prosthetic users in rehabilitation using accelerometry: Validation study},
  journal = {Journal of Rehabilitation and Assistive Technologies Engineering},
  year = {2016},
  doi = {10.1177/2055668316642387}
}
Kostenlos starten →