EUVIMEDEUVIMED entdecken
Open Access

Intelligent wearable devices with audio collection capabilities to assess chronic obstructive pulmonary disease severity

Intelligente Wearables mit Audioerfassung zur Beurteilung der COPD-Schwere

Chunbo Zhang iD, Kunyao Yu, Zhe Jin, Yingcong Bao, Cheng Zhang, Jiping Liao, Guangfa Wang iD

DIGITAL HEALTH · 2025-02 · 11

DOI: 10.1177/20552076251320730 ↗

Auf einen Blick

Bei 29 stabilen COPD-Patient:innen korrelierten Audio- und Vitaldaten aus einer Smartwatch mit der GOLD-Klassifikation; Hustenlaute zeigten die stärkste Korrelation.

Fragestellung

Wie gut lässt sich die COPD-Schwere mit physiologischen Parametern und Audioaufnahmen einer Smartwatch einschätzen?

Methode

29 COPD-Patient:innen; Lungenfunktion, EKG, Blutgase und 6-Minuten-Gehtest; 7–14 Tage Smartwatch-Monitoring sowie 2-mal täglich Husten- und Blaslaute; Korrelations- und Regressionsanalyse.

Ergebnisse

HR, HRV, RR, Husten- und Blaslaute waren signifikant mit dem GOLD-Grad korreliert. Hustenlaute: r = 0,7617; p < .001. Das Modell mit RR und Blaslauten erreichte 0,77–0,89 Genauigkeit; Cohen's kappa 0,6757.

Schlussfolgerung der Autor:innen

Audio-Daten korrelierten bei COPD stärker mit der Lungenfunktion als physiologische Parameter. Eine Smartwatch mit Audioerfassung bewertete die COPD-Schwere wirksam.

Kernaussagen

  • Hustenlaute hatten die stärkste Korrelation mit dem GOLD-Grad.
  • RR und Blaslaute waren Schlüsselprädiktoren im Modell.
  • Audio-Daten korrelierten stärker mit der Lungenfunktion als Vitaldaten.

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Fachartikel in der Zeitschrift DIGITAL HEALTH, online erschienen am 13.03.2025. Verfasst von Chunbo Zhang und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Respiratory and Critical Care Medicine, Peking University First Hospital, Beijing, China. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Beijing Health Technologies Promotion Program, the Capital’s Funds for Health Improvement and Research, National High Level Hospital Clinical Research Funding.

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Intelligent wearable devices have potential for chronic obstructive pulmonary disease (COPD) monitoring, but the effectiveness of combining cough and blowing sounds for disease assessment is unclear.

Objective

The objective was to assess COPD severity via physiological parameters and audio data collected by a smartwatch.

Methods

COPD patients underwent lung function tests, electrocardiograms, blood gas analysis, and 6-min walk tests. The patients’ peripheral arterial oxygen saturation (SpO 2 ), heart rate variability (HRV), heart rate (HR), and respiratory rate (RR) were continuously monitored via a smartwatch for 7–14 days, and voluntary cough and forceful blowing sounds were recorded twice daily. The HR, SpO 2 , and RR were categorized into all-day, sleep, and wake periods and summarized using the mean, standard deviation, median, 25th percentile, 75th percentile and percent variation. The correlations among lung function, physiological parameters, and audio data were analyzed to develop a model for predicting COPD severity.

Results

Twenty-nine stable patients, with a mean age of 67.0 ± 5.8 years, were enrolled, and 89.7% were male. HR, HRV, RR, cough sounds, and blowing sounds were significantly correlated with the Global Initiative for Chronic Obstructive Lung Disease (GOLD) grade, with cough sounds showing the highest correlation (r = 0.7617, p < .001). Cough sounds also had the strongest correlation with the mean 6-minute walking distance (r = 0.6847, p < .001), whereas blowing sounds had the strongest correlation with the Body mass index, airflow Obstruction, Dyspnea, and Exercise capacity index (r = −0.6749, p < .001). A logistic regression model using the RR and blowing sounds as key predictors achieved accuracies of 0.77–0.89 in determining the GOLD grade, with a Cohen's kappa coefficient of 0.6757.

Conclusions

Audio data were more strongly correlated with lung function in COPD patients than were physiological parameters. A smartwatch with audio collection capabilities effectively assessed COPD severity. Trial Registration ClinicalTrials.gov NCT05551169

Schlagwörter

Stichwörter der Autor:innen

COPDWearable devicesdisease diagnosispulmonary functiontelehealth

Autor:innen und Institutionen

  • Department of Respiratory and Critical Care Medicine, Peking University First Hospital, Beijing, China.

ORCID-Profile: Chunbo Zhang, Guangfa Wang

Förderung

  • Beijing Health Technologies Promotion Program
  • the Capital’s Funds for Health Improvement and Research
  • National High Level Hospital Clinical Research Funding

Zitieren

APA

Zhang, C., Yu, K., Jin, Z., Bao, Y., Zhang, C., Liao, J., & Wang, G. (2025). Intelligent wearable devices with audio collection capabilities to assess chronic obstructive pulmonary disease severity. DIGITAL HEALTH, 11. https://doi.org/10.1177/20552076251320730

Vancouver

Zhang C, Yu K, Jin Z, Bao Y, Zhang C, Liao J, et al. Intelligent wearable devices with audio collection capabilities to assess chronic obstructive pulmonary disease severity. DIGITAL HEALTH. 2025;11. doi:10.1177/20552076251320730

BibTeX
@article{Zhang2025,
  author = {Zhang, Chunbo and Yu, Kunyao and Jin, Zhe and Bao, Yingcong and Zhang, Cheng and Liao, Jiping and Wang, Guangfa},
  title = {Intelligent wearable devices with audio collection capabilities to assess chronic obstructive pulmonary disease severity},
  journal = {DIGITAL HEALTH},
  year = {2025},
  volume = {11},
  doi = {10.1177/20552076251320730}
}
Kostenlos starten →