Kontrollierte klinische Studie in der Zeitschrift American Journal of Rhinology & Allergy, 2025. Verfasst von Jae-Min Shin und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otorhinolaryngology - Head and Neck Surgery, College of Medicine, Korea University, Seoul, South Korea. Hauptthemen laut MEDLINE: COVID-19, COVID-19 Testing, Nasopharynx, Robotics. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 23 Literaturangaben. Gefördert von: Ministry of Science and ICT, South Korea, Technology Development Program, Korea University Medical Center and Anam Hospital.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To meet the high demand for polymerase chain reaction (PCR) tests to diagnose COVID-19 and rapidly control the outbreak, an efficient and safe molecular diagnostic protocol is necessary. In this study, we evaluated the efficacy and safety of the medical robot developed for non-face-to-face nasopharyngeal swab specimen collection.

Methods

In a nonclinical study, an otorhinolaryngologist collected swab specimens manually and using a medical robot. In a single-institution, randomized, open-label, prospective, exploratory clinical trial, nasopharyngeal swab specimens were collected from the enrolled participants both manually and by using the medical robot.

Results

Evaluation of the efficacy and safety of nasopharyngeal swab collection using a medical robot was assessed. After the operation of the robot, subjective discomfort experienced by the participants and any side effects or abnormalities in the nose were also monitored. Preliminary nonclinical data revealed comparable results between robotic and manual methods in terms of RNA metrics and cytokeratin-8 expression. Minor initial damage to A549 cells by the robot improved with subsequent use. In the clinical setting, the robot-assisted technique yielded a 92.31% detection rate for human RNase P, while the manual method achieved 100%. Post-swabbing discomfort reported by participants was similar for both methods and resolved within 48 h.

Conclusions

The medical robot system could efficiently, safely, and accurately collect nasopharyngeal swab samples in a non-face-to-face manner. Its installation in respiratory clinics, airports, or ports could minimize the infection risk between individuals and healthcare workers, thereby contributing to an efficient distribution of medical resources.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

COVID-19COVID-19 TestingNasopharynxRoboticsSARS-CoV-2Specimen HandlingAdultAgedFemaleHumansMaleMiddle AgedProspective Studies

Stichwörter der Autor:innen

clinical trialmedical robotnasopharyngeal swab

Autor:innen und Institutionen

  • Department of Otorhinolaryngology - Head and Neck Surgery, College of Medicine, Korea University, Seoul, South Korea.
  • Mucosal Immunology Institute, College of Medicine, Korea University, Seoul, South Korea.
  • Department of Medical Assistant Robot, Korea Institute of Machinery and Materials, Daegu, South Korea.
  • Biot Korea Inc., Gwangju, South Korea.

ORCID-Profile: Jae-Min Shin, Munsoo Han

Förderung

  • Ministry of Science and ICT, South Korea
  • Technology Development Program
  • Korea University Medical Center and Anam Hospital
  • National Research Foundation of Korea
  • Korea Health Industry Development Institute

Zitieren

APA

Shin, J. M., Han, M., Lee, D., Seo, J., Lee, J. M., Chang, Y., & Kim, T. H. (2025). Efficacy and Safety of a Medical Robot for Non-Face-to-Face Nasopharyngeal Swab Specimen Collection: Nonclinical and Clinical Trial Findings for COVID-19 Testing. American Journal of Rhinology & Allergy. https://doi.org/10.1177/19458924251323363

Vancouver

Shin JM, Han M, Lee D, Seo J, Lee JM, Chang Y, et al. Efficacy and Safety of a Medical Robot for Non-Face-to-Face Nasopharyngeal Swab Specimen Collection: Nonclinical and Clinical Trial Findings for COVID-19 Testing. American Journal of Rhinology & Allergy. 2025. doi:10.1177/19458924251323363

BibTeX
@article{Shin2025,
  author = {Shin, Jae-Min and Han, Munsoo and Lee, Dabin and Seo, Joonho and Lee, Jeoung-Min and Chang, Yeongjun and Kim, Tae Hoon},
  title = {Efficacy and Safety of a Medical Robot for Non-Face-to-Face Nasopharyngeal Swab Specimen Collection: Nonclinical and Clinical Trial Findings for COVID-19 Testing},
  journal = {American Journal of Rhinology & Allergy},
  year = {2025},
  doi = {10.1177/19458924251323363}
}
Kostenlos starten →