Feasibility and Performance of Continuous Glucose Monitoring to Guide Computerized Insulin Infusion Therapy in Cardiovascular Intensive Care Unit
Journal of Diabetes Science and Technology · 2024-05 · 18 (3) · 562-569
DOI: 10.1177/19322968241241005 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Diabetes Science and Technology, online erschienen am 02.04.2024. Verfasst von Lynn Ang und 14 weiteren Autor:innen. Beteiligte Einrichtung: Division of Metabolism, Endocrinology & Diabetes, Department of Internal Medicine, University of Michigan, Ann Arbor, MI, USA. Hauptthemen laut MEDLINE: Continuous Glucose Monitoring, Diabetes Mellitus, Hypoglycemic Agents, Insulin. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIDDK NIH HHS, JDRF, National Institute of Diabetes and Digestive and Kidney Diseases.
Themen: Intensivpflege
Abstract (Originalsprache)
Background
We evaluated the feasibility of real-time continuous glucose monitoring (CGM) for titrating continuous intravenous insulin infusion (CII) to manage hyperglycemia in postoperative individuals in the cardiovascular intensive care unit and assessed their accuracy, nursing acceptance, and postoperative individual satisfaction.
Methods
Dexcom G6 CGM devices were applied to 59 postsurgical patients with hyperglycemia receiving CII. A hybrid approach combining CGM with periodic point-of-care blood glucose (POC-BG) tests with two phases (initial-ongoing) of validation was used to determine CGM accuracy. Mean and median absolute relative differences and Clarke Error Grid were plotted to evaluate the CGM accuracy. Surveys of nurses and patients on the use of CGMs experience were conducted and results were analyzed.
Results
In this cohort (mean age 64, 32% female, 32% with diabetes) with 864 paired POC-BG and CGM values analyzed, mean and median absolute relative difference between POC-BG and CGM values were 13.2% and 9.8%, respectively. 99.7% of paired CGM and POC-BG were in Zones A and B of the Clarke Error Grid. Responses from nurses reported CGMs being very or quite convenient (n = 28; 93%) and it was favored over POC-BG testing (n = 28; 93%). Majority of patients (n = 42; 93%) reported their care process using CGM as being good or very good.
Conclusion
This pilot study demonstrates the feasibility, accuracy, and nursing convenience of adopting CGM via a hybrid approach for insulin titration in postoperative settings. These findings provide robust rationale for larger confirmatory studies to evaluate the benefit of CGM in postoperative care to improve workflow, enhance health outcomes, and cost-effectiveness.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Division of Metabolism, Endocrinology & Diabetes, Department of Internal Medicine, University of Michigan, Ann Arbor, MI, USA.
- Department of Pathology, University of Michigan, Ann Arbor, MI, USA.
- Department of Biostatistics, University of Michigan, Ann Arbor, MI, USA.
- Department of Pharmacy, University of Michigan, Ann Arbor, MI, USA.
- Division of Endocrinology and Metabolism, Department of Medicine, Michigan State University, East Lansing, MI, USA.
- Michigan Department of Nursing, University of Michigan, Ann Arbor, MI, USA.
ORCID-Profile: Lynn Ang, Yu Kuei Lin, Rodica Pop-Busui, Roma Gianchandani
Förderung
- NIDDK NIH HHS
- JDRF
- National Institute of Diabetes and Digestive and Kidney Diseases
Zitieren
APA
Ang, L., Lin, Y. K., Schroeder, L. F., Huang, Y., DeGeorge, C. A., Arnold, P., Akanbi, F., Knotts, S., DuBois, E., Desbrough, N., Qu, Y., Freeman, R., Esfandiari, N. H., Pop-Busui, R., & Gianchandani, R. (2024). Feasibility and Performance of Continuous Glucose Monitoring to Guide Computerized Insulin Infusion Therapy in Cardiovascular Intensive Care Unit. Journal of Diabetes Science and Technology, 18(3), 562-569. https://doi.org/10.1177/19322968241241005
Vancouver
Ang L, Lin YK, Schroeder LF, Huang Y, DeGeorge CA, Arnold P, et al. Feasibility and Performance of Continuous Glucose Monitoring to Guide Computerized Insulin Infusion Therapy in Cardiovascular Intensive Care Unit. Journal of Diabetes Science and Technology. 2024;18(3):562-569. doi:10.1177/19322968241241005
BibTeX
@article{Ang2024,
author = {Ang, Lynn and Lin, Yu Kuei and Schroeder, Lee F and Huang, Yiyuan and DeGeorge, Christina A and Arnold, Patrick and Akanbi, Folake and Knotts, Sharon and DuBois, Elizabeth and Desbrough, Nicole and Qu, Yunyan and Freeman, Regi and Esfandiari, Nazanene H and Pop-Busui, Rodica and Gianchandani, Roma},
title = {Feasibility and Performance of Continuous Glucose Monitoring to Guide Computerized Insulin Infusion Therapy in Cardiovascular Intensive Care Unit},
journal = {Journal of Diabetes Science and Technology},
year = {2024},
volume = {18},
number = {3},
pages = {562--569},
doi = {10.1177/19322968241241005}
}
EUVIMED