Fachartikel in der Zeitschrift Shoulder & Elbow, online erschienen am 13.04.2026. Verfasst von Erel Ben-Ari und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, Division of Shoulder and Elbow, NYU Langone Health, New York, NY, USA. Die Arbeit nennt 47 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Three-dimensional (3D) preoperative planning and intraoperative computer-assisted navigation (CAN) may influence baseplate augmentation and screw fixation in reverse total shoulder arthroplasty (RTSA). This study evaluated how these technologies affect augmented baseplate use and screw length. This is a Level IV retrospective study.

Methods

Three hundred primary RTSAs for glenohumeral degenerative conditions (2014–2024) were reviewed. Patients were divided into three groups: (1) 2D-CT planning without CAN (2014–2017, n = 100), (2) early 3D-CT with CAN (2019–2022, n = 100), and (3) recent 3D-CT with CAN (2022–2024, n = 100). Demographics, glenoid morphology, baseplate type, and screw characteristics were compared.

Results

Augmented baseplate use was higher in the 3D-CT groups versus 2D-CT (40% and 43% vs. 15%; P < .01) with no demographic or glenoid morphology differences (P > .05). CAN was associated with longer screws when four screws were used (24.6 mm and 25.3 mm vs. 20.8 mm; P < .001) and when comparing the two longest screws (28.5 mm and 27.3 mm vs. 22.9 mm; P < .01). Screw length did not differ between augmented and standard baseplates.

Discussion

3D planning and CAN increased augmented baseplate use and screw length. Equivalent screw lengths between augmented and standard baseplates suggest that augmentation does not compromise fixation, supporting its safe use in appropriately selected patients.

Schlagwörter

Stichwörter der Autor:innen

Reverse total shoulder arthroplastybaseplate augmentationcomputer-assisted navigation (CAN)glenoid deformityscrew lengththree-dimensional (3D) planning

Autor:innen und Institutionen

  • Department of Orthopaedic Surgery, Division of Shoulder and Elbow, NYU Langone Health, New York, NY, USA.
  • Division of Orthopaedic Surgery, Tel-Hashomer "Sheba" Medical Center, Sackler School of Medicine, Tel-Aviv University, Ramat Gan, Israel.
  • Department of Orthopaedic Surgery, Rabin Medical Center, Petach Tikva, Israel.
  • Sackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel.

ORCID-Profile: Erel Ben-Ari

Zitieren

APA

Ben-Ari, E., Rettig, S. A., Rosenthal, Y., Virk, M. S., & Zuckerman, J. D. (2026). The impact of three-dimensional planning and navigation on augmented baseplate use and screw length in reverse shoulder arthroplasty. Shoulder & Elbow. https://doi.org/10.1177/17585732261433989

Vancouver

Ben-Ari E, Rettig SA, Rosenthal Y, Virk MS, Zuckerman JD. The impact of three-dimensional planning and navigation on augmented baseplate use and screw length in reverse shoulder arthroplasty. Shoulder & Elbow. 2026. doi:10.1177/17585732261433989

BibTeX
@article{BenAri2026,
  author = {Ben-Ari, Erel and Rettig, Samantha A and Rosenthal, Yoav and Virk, Mandeep S and Zuckerman, Joseph D},
  title = {The impact of three-dimensional planning and navigation on augmented baseplate use and screw length in reverse shoulder arthroplasty},
  journal = {Shoulder & Elbow},
  year = {2026},
  doi = {10.1177/17585732261433989}
}
Kostenlos starten →