Current utility, instruments, and future directions for intra-renal pressure management during ureteroscopy: scoping review by global research in intra-renal pressure collaborative group initiative
Therapeutic Advances in Urology · 2025-01 · 17
DOI: 10.1177/17562872251314809 ↗Kurz gefasst
Scoping Review in der Zeitschrift Therapeutic Advances in Urology, online erschienen am 31.01.2025. Verfasst von Steffi Kar Kei Yuen und 12 weiteren Autor:innen. Beteiligte Einrichtung: SH Ho Urology Centre, Department of Surgery, The Chinese University of Hong Kong, Hong Kong, China. Die Arbeit wurde bisher 44-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background
Technical advancements and intrarenal pressure are synergistic in improving perioperative outcomes during flexible ureteroscopy (FURS). Mismanaged intra-renal pressure (IRP) has negative consequences and it is associated with an increased risk of sepsis, bleeding, pelvicalyceal fluid extravasation, and even collecting system injuries and acute as well as chronic renal failure. The cornerstone of a safe FURS is the ability to continuously monitor IRP to avoid elevation of IRP above the normal range of 10 mmHg.
Objectives
This scoping review aims to report the current state of real-time IRP monitoring in in vivo clinical studies and the various monitoring methods and technology to understand how this may be best used in daily clinical practice. Eligibility criteria: A systematic literature search was conducted. Only in vivo clinical studies published in English documenting IRP measurement methodologies during semirigid or flexible ureteroscopy for urolithiasis management were included.
Results
Out of 1326 retrieved papers, 17 studies met the inclusion criteria, comprising 2 randomized controlled trials, 2 retrospective studies, and 13 observational studies. Current noninvasive IRP monitoring devices include ureteric catheters placed retrogradely or via percutaneous tubing and connected to pressure transducers, amplified by cardiology-used pressure sensing systems or urodynamic systems, automated pressure-regulating systems, pressure sensing guidewires, and IRP sensing flexible ureteroscopes. The review revealed significant variations in IRP measurement methods, reporting units, and irrigation techniques. Notably, elevated IRP above 30 mmHg was consistently associated with increased postoperative complications, including sepsis.
Conclusion
Current clinical studies have only ascertained that increased IRP above 40 mmHg positively correlates with infectious complications and postoperative pain. No standardized values are available to predefine safe thresholds in practice. With the availability of noninvasive tools for IRP monitoring, future research should focus on multicenter studies to establish reference ranges and best practices for IRP management, ultimately improving patient outcomes in endourological procedures.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- SH Ho Urology Centre, Department of Surgery, The Chinese University of Hong Kong, Hong Kong, China.
- Department of Urology, Minimally Invasive Surgery Center, The First Affiliated Hospital of Guangzhou Medical University, Guangzhou, Guangdong, China.
- Department of Surgery, Tuen Mun Hospital, Tuen Mun, Hong Kong, China.
- Urology Unit, Azienda Ospedaliero-Universitaria delle Marche, Ancona, Italy.
- Division of Urology, University of Montreal Hospital Center, Montreal, QC, Canada.
- Department of Urology, Rainbow Hospital, S.N. Medical College, Agra, UP, India.
ORCID-Profile: Steffi Kar Kei Yuen, Daniele Castellani, Bhaskar Somani, Vineet Gauhar
Zitieren
APA
Yuen, S. K. K., Zhong, W., Chan, Y. S., Castellani, D., Bhojani, N., Agarwal, M. S., Tokas, T., Croghan, S., Jung, H., Herrmann, T., Somani, B., Gauhar, V., & Global Research on Intra-renal Pressure Collaborative Group (2025). Current utility, instruments, and future directions for intra-renal pressure management during ureteroscopy: scoping review by global research in intra-renal pressure collaborative group initiative. Therapeutic Advances in Urology, 17. https://doi.org/10.1177/17562872251314809
Vancouver
Yuen SKK, Zhong W, Chan YS, Castellani D, Bhojani N, Agarwal MS, et al. Current utility, instruments, and future directions for intra-renal pressure management during ureteroscopy: scoping review by global research in intra-renal pressure collaborative group initiative. Therapeutic Advances in Urology. 2025;17. doi:10.1177/17562872251314809
BibTeX
@article{Yuen2025,
author = {Yuen, Steffi Kar Kei and Zhong, Wen and Chan, Yun Sang and Castellani, Daniele and Bhojani, Naeem and Agarwal, Madhu Sudan and Tokas, Theodoros and Croghan, Stefanie and Jung, Helene and Herrmann, Thomas and Somani, Bhaskar and Gauhar, Vineet and Global Research on Intra-renal Pressure Collaborative Group, Global Research on Intra-renal Pressure Collaborative Group},
title = {Current utility, instruments, and future directions for intra-renal pressure management during ureteroscopy: scoping review by global research in intra-renal pressure collaborative group initiative},
journal = {Therapeutic Advances in Urology},
year = {2025},
volume = {17},
doi = {10.1177/17562872251314809}
}