Long-term results of the Aptis TM total distal radioulnar joint prosthesis after previous failed surgical procedures
Journal of Hand Surgery (European Volume) · 2024
DOI: 10.1177/17531934231192375 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Hand Surgery (European Volume), 2024. Verfasst von Gregorio Martínez Villén und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic and Traumatology (Hand and Reconstructive Surgery Unit), Miguel Servet University Hospital, Zaragoza, Spain. Hauptthemen laut MEDLINE: Joint Prosthesis, Arthroplasty, Replacement. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 32 Literaturangaben.
Abstract (Originalsprache)
We present a prospective study with the results of ten Aptis total distal radioulnar joint replacements in patients who had one to five previous operations. The mean postoperative follow-up was 9.7 years (range 3–14.7). The mean postoperative range of motion of the wrist improved by 9° for extension, 15° for flexion, 10° for pronation and 14° for supination, achieving values equivalent to 88%, 78%, 97% and 88% of those of the non-operated hand. There were minimal changes in radial and ulnar deviations. Grip strength improved by 7 kg, pain decreased by 8 points, QuickDASH and Mayo wrist scores improved by 51 and 53 points, respectively. Two implants had radiolucency less than 1 mm. Serum titanium ion levels were slightly elevated in two patients. Three prostheses required revision surgery for heterotopic ossification, a prominent radial screw and a periprosthetic fracture. No prostheses were removed. Six patients returned to their former professional activities and four patients adapted their jobs. Level of evidence: IV
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopaedic and Traumatology (Hand and Reconstructive Surgery Unit), Miguel Servet University Hospital, Zaragoza, Spain.
- Department of Orthopaedic and Traumatology, University Hospital Royo Villanova, Zaragoza, Spain.
- Department of Clinical Biochemistry, Miguel Servet University Hospital, Zaragoza, Spain.
ORCID-Profile: Gregorio Martínez Villén
Zitieren
APA
Martínez Villén, G., Espada Blasco, C., Rodríguez Nogué, L., & García González, E. (2024). Long-term results of the Aptis TM total distal radioulnar joint prosthesis after previous failed surgical procedures. Journal of Hand Surgery (European Volume). https://doi.org/10.1177/17531934231192375
Vancouver
Martínez Villén G, Espada Blasco C, Rodríguez Nogué L, García González E. Long-term results of the Aptis TM total distal radioulnar joint prosthesis after previous failed surgical procedures. Journal of Hand Surgery (European Volume). 2024. doi:10.1177/17531934231192375
BibTeX
@article{MartnezVilln2024,
author = {Martínez Villén, Gregorio and Espada Blasco, Carlos and Rodríguez Nogué, Luis and García González, Elena},
title = {Long-term results of the Aptis TM total distal radioulnar joint prosthesis after previous failed surgical procedures},
journal = {Journal of Hand Surgery (European Volume)},
year = {2024},
doi = {10.1177/17531934231192375}
}
EUVIMED