Injury of the superficial femoral artery and posterior tibial artery associated with an atherectomy device
Vascular · 2025
DOI: 10.1177/17085381241276608 ↗Kurz gefasst
Fallbericht in der Zeitschrift Vascular, 2025. Verfasst von Ahmed M Afifi und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Surgery, University of Toledo Medical Center, Toledo, OH, USA. Hauptthemen laut MEDLINE: Femoral Artery, Tibial Arteries, Atherectomy, Peripheral Arterial Disease. Die Arbeit nennt 10 Literaturangaben.
Abstract (Originalsprache)
Objectives
Rotational atherectomy can offer a viable treatment for occlusive peripheral artery disease; maintaining the minimal invasiveness of an endovascular procedure, while allowing for a more complete lesion debridement compared with balloon angioplasty. This case report outlines a complication of guidewire entrapment associated with rotational atherectomy in the superficial femoral artery (SFA).
Methods
A 57-year-old male underwent an atherectomy with Rotorex for left lower limb foot pain. During the procedure, the guidewire was suctioned into the atherectomy device, preventing any further advancement of the device and damaging the SFA and posterior tibial artery (PTA).
Results
The atherectomy device was withdrawn and a new vascular access site was gained in the left PTA. A covered stent was inserted to treat the original SFA lesion, and balloon angioplasty was used to repair the device-induced damaged to the PTA.
Conclusion
While guidewire complications have been previously reported, this case report details the first reported case, to our knowledge, of guidewire entrapment while using a rotational atherectomy device. Knowledge of this possible complication of rotational atherectomy can aid in clinical decision making when choosing between treatments for peripheral vascular disease.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Surgery, University of Toledo Medical Center, Toledo, OH, USA.
ORCID-Profile: Samuel C Dunseith
Zitieren
APA
Afifi, A. M., Dunseith, S. C., & Nazzal, M. (2025). Injury of the superficial femoral artery and posterior tibial artery associated with an atherectomy device. Vascular. https://doi.org/10.1177/17085381241276608
Vancouver
Afifi AM, Dunseith SC, Nazzal M. Injury of the superficial femoral artery and posterior tibial artery associated with an atherectomy device. Vascular. 2025. doi:10.1177/17085381241276608
BibTeX
@article{Afifi2025,
author = {Afifi, Ahmed M and Dunseith, Samuel C and Nazzal, Munier},
title = {Injury of the superficial femoral artery and posterior tibial artery associated with an atherectomy device},
journal = {Vascular},
year = {2025},
doi = {10.1177/17085381241276608}
}