Evaluating the between-day reliability and likelihood of change of a test battery incorporating vastus lateralis muscle thickness, ankle-brachial pressure index, maximal voluntary torque, and six-minute walk test in patients with claudication

Thomas Parkington iD, David Broom, Thomas Maden-Wilkinson, Shah Nawaz iD, Markos Klonizakis

Vascular · 2025

DOI: 10.1177/17085381241257735 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Vascular, 2025. Verfasst von Thomas Parkington und 4 weiteren Autor:innen. Beteiligte Einrichtung: Physical Activity, Wellness and Public Health Research Group, School of Sport and Physical Activity, Sheffield Hallam University, Sheffield, UK. Hauptthemen laut MEDLINE: Intermittent Claudication, Ankle Brachial Index, Walk Test, Quadriceps Muscle. Die Arbeit nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

The study aims to evaluate the between-day reliability of a proposed test battery for patients with claudication that can be used for monitoring the effectiveness of exercise interventions and other therapeutic strategies tailored to this patient population.

Methods

Twenty-five men with claudication were recruited. The test battery consisted of the Vastus Lateralis muscle thickness (VL-MT), ankle-brachial pressure index (ABI), unilateral isometric knee extension maximal voluntary torque (MVT) and 6-minute walk test (6MWT). A single investigator conducted the tests for each patient on two separate testing sessions (T1 and T2) 5–7 days apart.

Results

Good to excellent reliability was observed for VL-MT (ICC = 0.95, 95% LOA = ±3.10 mm, SEM = 0.81 mm), ABI (ICC = 0.97, 95% LOA = ±0.10, SEM = 0.02), MVT (ICC = 0.97, 95% LOA = ±24.0 N·m, SEM = 6.31 N·m), 6MWT distance (ICC = 0.99, 95% LOA = ±39.6 m, SEM = 11.0 m), 6MWT time to claudication (ICC = 0.99, 95% LOA = ±30.8 s, SEM = 7.8 s), and 6MWT ratings of pain (ICC = 0.87, 95% LOA = ±2.4 CR-10 + , SEM = 0.7 CR-10 + ). Analysis derived from reliability data indicates a change of 1.4 mm for VL-MT, 0.14 for ABI, 12 N·m for MVT, 25 m for 6MWT distance, 15 s for 6MWT time to claudication and 1 CR-10 + for 6MWT ratings of pain is required to be interpreted as the minimum ‘likely’ change (76% chance).

Conclusions

The test battery provides a reliable assessment of patients with claudication and can be widely used to evaluate the effects of exercise programmes and other therapeutic interventions. For the individual, changes in VL-MT, ABI, MVT, and 6MWT greater than the minimum likely change as a result of an exercise programme or an intervention are likely changes and less influenced by error associated with the test.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Intermittent ClaudicationAnkle Brachial IndexWalk TestQuadriceps MusclePeripheral Arterial DiseaseExercise ToleranceMuscle StrengthHumansMaleReproducibility of ResultsPredictive Value of TestsMiddle AgedTime FactorsAgedTorqueIsometric Contraction

Stichwörter der Autor:innen

Ankle-brachial indexPeripheral artery diseaseSix-minute walk testmaximal voluntary torquemuscle thickness

Autor:innen und Institutionen

  • Physical Activity, Wellness and Public Health Research Group, School of Sport and Physical Activity, Sheffield Hallam University, Sheffield, UK.
  • Centre for Physical Activity, Sport and Exercise Sciences, Coventry University, Coventry, UK.
  • Sheffield Vascular Institute, Sheffield Teaching Hospitals NHS Foundation Trust, Sheffield, UK.
  • Lifestyle, Exercise and Nutrition Improvement Research Group, Department of Nursing and Midwifery, Sheffield Hallam University, Sheffield, UK.

ORCID-Profile: Thomas Parkington, Shah Nawaz

Zitieren

APA

Parkington, T., Broom, D., Maden-Wilkinson, T., Nawaz, S., & Klonizakis, M. (2025). Evaluating the between-day reliability and likelihood of change of a test battery incorporating vastus lateralis muscle thickness, ankle-brachial pressure index, maximal voluntary torque, and six-minute walk test in patients with claudication. Vascular. https://doi.org/10.1177/17085381241257735

Vancouver

Parkington T, Broom D, Maden-Wilkinson T, Nawaz S, Klonizakis M. Evaluating the between-day reliability and likelihood of change of a test battery incorporating vastus lateralis muscle thickness, ankle-brachial pressure index, maximal voluntary torque, and six-minute walk test in patients with claudication. Vascular. 2025. doi:10.1177/17085381241257735

BibTeX
@article{Parkington2025,
  author = {Parkington, Thomas and Broom, David and Maden-Wilkinson, Thomas and Nawaz, Shah and Klonizakis, Markos},
  title = {Evaluating the between-day reliability and likelihood of change of a test battery incorporating vastus lateralis muscle thickness, ankle-brachial pressure index, maximal voluntary torque, and six-minute walk test in patients with claudication},
  journal = {Vascular},
  year = {2025},
  doi = {10.1177/17085381241257735}
}
Kostenlos starten →