Periprocedural thromboembolic complications of LEO baby stent in endovascular treatment of intracranial aneurysms: Experience in 149 patients
Interventional Neuroradiology · 2026
DOI: 10.1177/15910199231217547 ↗Kurz gefasst
Fachartikel in der Zeitschrift Interventional Neuroradiology, 2026. Verfasst von Yunan Shen und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Interventional Radiology, The First Affiliated Hospital of Nanjing Medical University, Nanjing, China. Hauptthemen laut MEDLINE: Intracranial Aneurysm, Stents, Thromboembolism, Embolization, Therapeutic. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Objective
To report periprocedural thromboembolic complications of LEO Baby stent-assisted coiling of wide-necked intracranial aneurysms and to analyze the possible influencing factors.
Methods
We retrospectively identified 149 patients with aneurysms who underwent LEO Baby stent-assisted embolization between October 2018 and March 2022. Clinical and radiographic data of patients were reviewed to determine whether a thromboembolic event had occurred. Multivariate logistic analysis was performed to identify significant factors associated with thromboembolic events.
Results
Successful stent deployment of the stent was achieved in all patients in the target artery. There were 66 patients (44.3%) with acutely ruptured aneurysms and 83 patients (55.7%) with unruptured aneurysms. Fourteen (9.4%, 95% confidence interval: 4.7%–14.1%) patients were confirmed to have developed a thromboembolic event, including nine patients with acute intraoperative thrombosis and five patients with postoperative thromboembolic events. The rate of thromboembolic events was 6.0% (5/83) in patients with unruptured aneurysms and 13.6% (9/66) in patients with acutely ruptured aneurysms. There was a trend toward an increased rate of thromboembolic events in patients with acute ruptured aneurysms ( p = 0.087). Thromboembolic events were significantly associated with the parent-artery diameter ( p = 0.010).
Conclusions
Our study demonstrates a low rate of thromboembolic complications in unruptured aneurysms treated with LEO Baby stent. Thromboembolic events appear to be more common in ruptured aneurysms. A small diameter of the parent artery is associated with an increased risk of thromboembolic complications, and more relevant studies are still needed.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Interventional Radiology, The First Affiliated Hospital of Nanjing Medical University, Nanjing, China.
ORCID-Profile: Heng Ni, Sheng Liu, Linbo Zhao
Zitieren
APA
Shen, Y., Ni, H., Jia, Z., Cao, Y., Lu, G., Liu, X., Wang, B., Zhou, C., Shi, H., Liu, S., & Zhao, L. (2026). Periprocedural thromboembolic complications of LEO baby stent in endovascular treatment of intracranial aneurysms: Experience in 149 patients. Interventional Neuroradiology. https://doi.org/10.1177/15910199231217547
Vancouver
Shen Y, Ni H, Jia Z, Cao Y, Lu G, Liu X, et al. Periprocedural thromboembolic complications of LEO baby stent in endovascular treatment of intracranial aneurysms: Experience in 149 patients. Interventional Neuroradiology. 2026. doi:10.1177/15910199231217547
BibTeX
@article{Shen2026,
author = {Shen, Yunan and Ni, Heng and Jia, Zhenyu and Cao, Yuezhou and Lu, Guangdong and Liu, Xinglong and Wang, Bin and Zhou, Chun and Shi, Haibin and Liu, Sheng and Zhao, Linbo},
title = {Periprocedural thromboembolic complications of LEO baby stent in endovascular treatment of intracranial aneurysms: Experience in 149 patients},
journal = {Interventional Neuroradiology},
year = {2026},
doi = {10.1177/15910199231217547}
}