EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY 4.0

Virtual implantation of Derivo Embolization Devices using PreSize Neurovascular software: Accuracy evaluation and size selection comparison study

Ngoc T Ngo iD, Fabian Flottmann, Francesco Iori, Mirko Bonfanti, Aikaterini Mandaltsi, Jens Fiehler, Maxim Bester

Interventional Neuroradiology · 2026

DOI: 10.1177/15910199231202272 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Interventional Neuroradiology, 2026. Verfasst von Ngoc T Ngo und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic and Interventional Neuroradiology, University Medical Center Hamburg-Eppendorf, Hamburg, Germany. Hauptthemen laut MEDLINE: Embolization, Therapeutic, Intracranial Aneurysm, Software, Stents. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Innovate UK.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Evaluating the deployed length of flow diverting stents (FDs) to select the optimal device size remains a challenging, yet crucial, task in aneurysm treatment. This study reports on the accuracy of PreSize Neurovascular (Oxford Heartbeat Ltd), a visualization and simulation software for FD intervention planning, in predicting FD deployed length, and on its impact on device size selection.

Methods

Imaging data from consecutive patients treated with Derivo Embolization Device (Acandis GmbH) were collected from University Medical Center Hamburg-Eppendorf and retrospectively analyzed. Accuracy evaluation: prediction accuracy was calculated by comparing deployed FD lengths measured from imaging data and simulated by PreSize. Size selection comparison: two Interventional Neuroradiologists (INR1, INR2), blinded to the devices deployed, used PreSize to select the optimal device size (diameter and length). Their choices were compared against the deployed devices selected by conventional planning.

Results

Among 98 implanted devices, PreSize predicted deployed FD length with a mean accuracy of 94.54% (95% confidence interval [93.72%, 95.35%]). Among 98 aneurysm cases, PreSize-informed device lengths were significantly shorter (Wilcoxon signed-rank test, INR1: W = 394, P < .001, INR2: W = 305, P < .001) by 4.13 and 4.18 mm on average, and up to 20 and 25 mm, for INR1 and INR2, respectively, than the conventionally selected FDs. In 32% of cases, PreSize-informed devices resulted in fewer vessel bends covered by the FD while achieving sufficient aneurysm coverage.

Conclusions

PreSize retrospectively predicted deployed FD lengths with high accuracy. Moreover, INRs in this study were more inclined to select shorter stent length in the simulation than they would have done conventionally.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Embolization, TherapeuticIntracranial AneurysmSoftwareStentsHumansRetrospective StudiesFemaleCerebral AngiographyMaleAgedMiddle Aged

Stichwörter der Autor:innen

Device selectionaneurysm treatmentflow diverter sizingintervention planningsimulation software

Autor:innen und Institutionen

  • Department of Diagnostic and Interventional Neuroradiology, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
  • Oxford Heartbeat Ltd, London, UK.

ORCID-Profile: Ngoc T Ngo

Förderung

  • Innovate UK

Zitieren

APA

Ngo, N. T., Flottmann, F., Iori, F., Bonfanti, M., Mandaltsi, A., Fiehler, J., & Bester, M. (2026). Virtual implantation of Derivo Embolization Devices using PreSize Neurovascular software: Accuracy evaluation and size selection comparison study. Interventional Neuroradiology. https://doi.org/10.1177/15910199231202272

Vancouver

Ngo NT, Flottmann F, Iori F, Bonfanti M, Mandaltsi A, Fiehler J, et al. Virtual implantation of Derivo Embolization Devices using PreSize Neurovascular software: Accuracy evaluation and size selection comparison study. Interventional Neuroradiology. 2026. doi:10.1177/15910199231202272

BibTeX
@article{Ngo2026,
  author = {Ngo, Ngoc T and Flottmann, Fabian and Iori, Francesco and Bonfanti, Mirko and Mandaltsi, Aikaterini and Fiehler, Jens and Bester, Maxim},
  title = {Virtual implantation of Derivo Embolization Devices using PreSize Neurovascular software: Accuracy evaluation and size selection comparison study},
  journal = {Interventional Neuroradiology},
  year = {2026},
  doi = {10.1177/15910199231202272}
}