Heterogeneous model to evaluate CFD in intracranial bifurcation aneurysms treated with the WEB device to predict angiographic outcome
Interventional Neuroradiology · 2025
DOI: 10.1177/15910199231152514 ↗Kurz gefasst
Fachartikel in der Zeitschrift Interventional Neuroradiology, 2025. Verfasst von Daniel Mantilla und 5 weiteren Autor:innen. Beteiligte Einrichtung: Interventional Radiology Department, Fundación oftalmológica de Santander - Clínica Ardila Lülle. Floridablanca, Colombia. Hauptthemen laut MEDLINE: Intracranial Aneurysm, Cerebral Angiography, Embolization, Therapeutic. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Introduction
The Woven EndoBridge device (WEB) was developed as an alternative to treat Wide-Necked bifurcation aneurysms. It has proven to be effective and safe, however, cases of recanalization have been reported. The purpose of this study was to evaluate and quantify hemodynamic parameters and indexes with CFD of the intracranial aneurysms before and after WEB simulation and to establish their relationship to complete occlusion. Materials and methods Using the heterogeneous model based on the marching cubes algorithm, we created 3D representations of 27 bifurcated intracranial aneurysms treated with the single-layer WEB device to evaluate hemodynamics parameters with CFD, calculated with and without the WEB.
Results
We observed a lower treatment entry concentration indices (ICI) (2.12 ± 1.31 versus 3.14 ± 0.93, p-value: 0.029) previous to placement of WEB and higher pre-treatment FN (7.56 ± 5.92 versus 3.35 ± 1.51, p-value: 0.018) and post-treatment FN (5.34 ± 5.89 versus 1.99 ± 0.83, p-value: 0.021) for cases with successful occlusions. Lower post-treatment SRa (197.81 ± 221.29 versus 80.02 ± 45.25, p-value: 0.044) and higher pre (0.11 ± 0.07 versus 0.25 ± 0.19, p-value: 0.011) and post-treatment MATT (0.69 ± 1.23 versus 1.02 ± 0.46, p-value: 0.006) were observed in non-occluded cases.
Conclusions
In our CFD analysis of the hemodynamic parameters of IA, we found lower ICI before the placement of the WEB device and higher FN pre- and post-treatment for cases with successful occlusions. Non-occluded cases had lower post-treatment SRa and higher pre-treatment and post-treatment MATT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Interventional Radiology Department, Fundación oftalmológica de Santander - Clínica Ardila Lülle. Floridablanca, Colombia.
- Interventional Radiology Department, Universidad Autónoma de Bucaramanga, Bucaramanga, Colombia.
- Faculté de Sciences, Université de Montpellier, Montpellier, France.
- Sim&Cure, Montpellier, France.
- Institut Montpelliérain Alexander, Grothendieck, CNRS, Univ. Montpellier, Montpellier, France.
- Radiology Department, Fundación oftalmológica de Santander, Clínica Ardila Lülle, Floridablanca, Colombia.
ORCID-Profile: Daniel Mantilla, Andrés F Ortiz, Daniela D Vera
Zitieren
APA
Mantilla, D., Berod, A., Ortiz, A. F., Vera, D. D., Nicoud, F., & Costalat, V. (2025). Heterogeneous model to evaluate CFD in intracranial bifurcation aneurysms treated with the WEB device to predict angiographic outcome. Interventional Neuroradiology. https://doi.org/10.1177/15910199231152514
Vancouver
Mantilla D, Berod A, Ortiz AF, Vera DD, Nicoud F, Costalat V. Heterogeneous model to evaluate CFD in intracranial bifurcation aneurysms treated with the WEB device to predict angiographic outcome. Interventional Neuroradiology. 2025. doi:10.1177/15910199231152514
BibTeX
@article{Mantilla2025,
author = {Mantilla, Daniel and Berod, Alain and Ortiz, Andrés F and Vera, Daniela D and Nicoud, Franck and Costalat, Vincent},
title = {Heterogeneous model to evaluate CFD in intracranial bifurcation aneurysms treated with the WEB device to predict angiographic outcome},
journal = {Interventional Neuroradiology},
year = {2025},
doi = {10.1177/15910199231152514}
}