Automated CT angiography collateral scoring in anterior large vessel occlusion stroke: A multireader study
Interventional Neuroradiology · 2025
DOI: 10.1177/15910199221150470 ↗Kurz gefasst
Fachartikel in der Zeitschrift Interventional Neuroradiology, 2025. Verfasst von Mohamed Sobhi Jabal und 16 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Mayo Clinic, Rochester, MN, USA. Hauptthemen laut MEDLINE: Computed Tomography Angiography, Collateral Circulation, Ischemic Stroke, Cerebral Angiography. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Themen: Schlaganfall
Abstract (Originalsprache)
Background
Computed tomography (CT) angiography collateral score (CTA-CS) is an important clinical outcome predictor following mechanical thrombectomy for ischemic stroke with large vessel occlusion (LVO). The present multireader study aimed to evaluate the performance of e-CTA software for automated assistance in CTA-CS scoring. Materials and Methods Brain CTA images of 56 patients with anterior LVO were retrospectively processed. Twelve readers of various clinical training, including junior neuroradiologists, senior neuroradiologists, and neurologists graded collateral flow using visual CTA-CS scale in two sessions separated by a washout period. Reference standard was the consensus of three expert readers. Duration of reading time, inter-rater reliability, and statistical comparison of readers’ performance metrics were analyzed between the e-CTA assisted and unassisted sessions. Results e-CTA assistance resulted in significant increase in mean accuracy (58.6% to 67.5%, p = 0.003), mean F1 score (0.574 to 0.676, p = 0.002), mean precision (58.8% to 68%, p = 0.007), and mean recall (58.7% to 69.9%, p = 0.002), especially with slight filling deficit (CTA-CS 2 and 3). Mean reading time was reduced across all readers (103.4 to 59.7 s, p = 0.001), and inter-rater agreement in CTA-CS assessment was increased (Krippendorff's alpha 0.366 to 0.676). Optimized occlusion laterality detection was also noted with mean accuracy (92.9% to 96.8%, p = 0.009).
Conclusion
Automated assistance for CTA-CS using e-CTA software provided helpful decision support for readers in terms of improving scoring accuracy and reading efficiency for physicians with a range of experience and training backgrounds and leading to significant improvements in inter-rater agreement.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Mayo Clinic, Rochester, MN, USA.
- Brainomix Limited, Oxford, UK.
- Oxford University Hospitals, NHS Foundation Trust, Oxford, UK.
- Department of Neurology, Mayo Clinic, Rochester, MN, USA.
ORCID-Profile: Mohamed Sobhi Jabal, David F Kallmes, George Harston, John Benson, Ajay Madhavan, Cem Bilgin
Zitieren
APA
Jabal, M. S., Kallmes, D. F., Harston, G., Campeau, N., Schwartz, K., Messina, S., Carr, C., Benson, J., Little, J., Nagelschneider, A., Madhavan, A., Nasr, D., Braksick, S., Klaas, J., Scharf, E., Bilgin, C., & Brinjikji, W. (2025). Automated CT angiography collateral scoring in anterior large vessel occlusion stroke: A multireader study. Interventional Neuroradiology. https://doi.org/10.1177/15910199221150470
Vancouver
Jabal MS, Kallmes DF, Harston G, Campeau N, Schwartz K, Messina S, et al. Automated CT angiography collateral scoring in anterior large vessel occlusion stroke: A multireader study. Interventional Neuroradiology. 2025. doi:10.1177/15910199221150470
BibTeX
@article{Jabal2025,
author = {Jabal, Mohamed Sobhi and Kallmes, David F and Harston, George and Campeau, Norbert and Schwartz, Kara and Messina, Steven and Carr, Carrie and Benson, John and Little, Jason and Nagelschneider, Alex and Madhavan, Ajay and Nasr, Deena and Braksick, Sherry and Klaas, James and Scharf, Eugene and Bilgin, Cem and Brinjikji, Waleed},
title = {Automated CT angiography collateral scoring in anterior large vessel occlusion stroke: A multireader study},
journal = {Interventional Neuroradiology},
year = {2025},
doi = {10.1177/15910199221150470}
}
EUVIMED