The effect of the size of the new contour neurovascular device for altering intraaneurysmal flow
Interventional Neuroradiology · 2025
DOI: 10.1177/15910199221145985 ↗Kurz gefasst
Fachartikel in der Zeitschrift Interventional Neuroradiology, 2025. Verfasst von Mariya S Pravdivtseva und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Neuroradiology, Section Biomedical Imaging, Molecular Imaging North Competence Center (MOIN CC), University Medical Cente…. Hauptthemen laut MEDLINE: Intracranial Aneurysm, Embolization, Therapeutic, Endovascular Procedures. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Zukunftsprogramm Wirtschaft of Schleswig-Holstein, Deutsche Forschungsgemeinschaft, Bundesministerium für Bildung und Forschung.
Abstract (Originalsprache)
Background
Recently, a novel intrasaccular device (contour neurovascular system, contour) was introduced to treat intracranial aneurysms. Contour is placed at thе aneurysm neck and reduces the intraaneurysmal blood inflow. Contour comes in a range of sizes to target different aneurysms. The efficiency of altering flow with contour and the effect of device size have not yet been investigated. Therefore, we studied the effect of the device size with patient-based aneurysm models using 2D digital subtraction angiography (DSA).
Methods
Three patient-based aneurysm models with necks ranging from 2.7 to 9.7 mm were produced, providing standardized testing conditions. Contours with diameters of 5, 11, and 14 mm were implanted into the models, four of each size. 2D DSA images were acquired before and after implanting contour (15 frames/s, manual contrast injection). After injecting angiographic contrast agent, the DSA signal was recorded over time to calculate the contrast washout time (WOT), which is a measure of flow diversion efficiency.
Results
All contour devices caused contrast agent stasis and increased WOT in aneurysm sac ( p -value = 0.0005). The median relative WOT was largest for 5-mm contour (6.6 ± 3.2) and similar for 11-mm contour (3.4 ± 2.6) and 14-mm contour (3.2 ± 3.8). The implantation procedure might affect WOT values even for contours of the same size; the overall relative WOT ranged between 1.5 and 10.89.
Conclusion
The 5-mm contour showed the longest WOT value in our study, while no apparent difference between 11-mm contour and 14-mm contour was found.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiology and Neuroradiology, Section Biomedical Imaging, Molecular Imaging North Competence Center (MOIN CC), University Medical Center Schleswig-Holstein (UKSH), Kiel University, Kiel, Germany.
- Department of Radiology and Neuroradiology, University Medical Center Schleswig-Holstein (UKSH), Kiel, Germany.
ORCID-Profile: Mariya S Pravdivtseva, Johannes Hensler, Fritz Wodarg
Förderung
- Zukunftsprogramm Wirtschaft of Schleswig-Holstein
- Deutsche Forschungsgemeinschaft
- Bundesministerium für Bildung und Forschung
Zitieren
APA
Pravdivtseva, M. S., Pravdivtsev, A. N., Peters, S., Hensler, J., Larsen, N., Hövener, J. B., Jansen, O., & Wodarg, F. (2025). The effect of the size of the new contour neurovascular device for altering intraaneurysmal flow. Interventional Neuroradiology. https://doi.org/10.1177/15910199221145985
Vancouver
Pravdivtseva MS, Pravdivtsev AN, Peters S, Hensler J, Larsen N, Hövener JB, et al. The effect of the size of the new contour neurovascular device for altering intraaneurysmal flow. Interventional Neuroradiology. 2025. doi:10.1177/15910199221145985
BibTeX
@article{Pravdivtseva2025,
author = {Pravdivtseva, Mariya S and Pravdivtsev, Andrey N and Peters, Sönke and Hensler, Johannes and Larsen, Naomi and Hövener, Jan-Bernd and Jansen, Olav and Wodarg, Fritz},
title = {The effect of the size of the new contour neurovascular device for altering intraaneurysmal flow},
journal = {Interventional Neuroradiology},
year = {2025},
doi = {10.1177/15910199221145985}
}
EUVIMED