Fallbericht in der Zeitschrift Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery, online erschienen am 06.08.2026. Verfasst von S Chris Malaisrie und 3 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiac Surgery, Bluhm Cardiovascular Institute, Northwestern University Feinberg School of Medicine, Chicago, IL, USA. Hauptthemen laut MEDLINE: Aortic Valve, Heart Valve Prosthesis, Heart Valve Prosthesis Implantation, Aortic Valve Stenosis. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 4 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objective
To demonstrate a supra-annular implantation of a composite valve-graft (CVG) constructed from a sutureless aortic valve within a Dacron graft for aortic root replacement. The anatomic rationale, preclinical validation, and clinical outcomes of this novel technique are described.
Methods
The anatomic rational for sizing of the valve and graft was examined by ex vivo implantation of a Perceval Plus XL valve (Corcym, Milan, Italy) within a 28 mm Valsalva Dacron graft (Terumo Aortic, Inchinnan, UK). Hydrodynamic in vitro testing with a ViVitro pulse duplicator system (ViVitro, Victoria, BC, Canada) validated the integrity of the CVG. The CVG was implanted supra-annular to the native aortic annulus in a clinical setting.
Results
A Perceval XL valve within a 28 mm Valsalva Dacron graft provided 8% oversize of the valve inflow ring within the graft collar. In vitro testing showed valve stability within the Dacron graft under physiologic conditions. The first SUpra-annular PERceval Bentall (SUPERB) was performed in a 75-year-old male patient with concomitant aortic stenosis and aortic root aneurysm. The CVG was constructed by (1) suturing the graft to the aortic annulus, (2) construction of coronary buttons, and (3) deployment of the sutureless valve with the Dacron graft in a supra-annular position.
Conclusions
SUPERB uses a Perceval XL valve within a 28 mm Valsalva Dacron graft for aortic root replacement. Positioning of the sutureless valve within the Dacron graft and supra-annular to the native aortic annulus decouples valve sizing from the restrictions of native aortic annular dimensions. This approach simplifies CVG sizing and potentially mitigates patient–prosthetic mismatch.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Division of Cardiac Surgery, Bluhm Cardiovascular Institute, Northwestern University Feinberg School of Medicine, Chicago, IL, USA.
- Department of Surgery, University of Ljubljana Faculty of Medicine, Slovenia.
- Chief Medical Officer, Corcym, Fort Collins, CO, USA.
- Tokyo D Tower Hospital, Japan.
ORCID-Profile: S Chris Malaisrie
Zitieren
APA
Malaisrie, S. C., Gersak, B., Whippo, B., & Snyder, M. (2026). Supra-Annular Perceval Bentall: From Concept to Patient. Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery, 21(4), 383-388. https://doi.org/10.1177/15569845261469838
Vancouver
Malaisrie SC, Gersak B, Whippo B, Snyder M. Supra-Annular Perceval Bentall: From Concept to Patient. Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery. 2026;21(4):383-388. doi:10.1177/15569845261469838
BibTeX
@article{Malaisrie2026,
author = {Malaisrie, S Chris and Gersak, Borut and Whippo, Beth and Snyder, Matthew},
title = {Supra-Annular Perceval Bentall: From Concept to Patient},
journal = {Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery},
year = {2026},
volume = {21},
number = {4},
pages = {383--388},
doi = {10.1177/15569845261469838}
}
EUVIMED