Physiologic Mitral Annuloplasty Ring Sizing: Surgical Technique and Animal Model Validation

Nikola Dobrilovic iD

Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery · 2026-03 · 21 (2) · 159-165

DOI: 10.1177/15569845261437227 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery, online erschienen am 11.05.2026. Verfasst von Nikola Dobrilovic. Beteiligte Einrichtung: Division of Cardiac Surgery, Department of Surgery, NorthShore University Hospital System, Chicago, IL, USA. Hauptthemen laut MEDLINE: Mitral Valve Annuloplasty, Mitral Valve, Heart Valve Prosthesis. Die Arbeit nennt 21 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

The current method for mitral annuloplasty ring sizing has remained essentially unchanged for a half century and offers surgeons limited actionable information. A novel technique is illustrated and validated in a cadaver swine heart model.

Methods

Twenty cadaver swine hearts were used to validate the novel sizing device and method. We hypothesized that the paired-ring sizer would force annuloplasty sutures to conform to its exact shape and therefore accurately predict changes in coaptation length when used to “downsize” the mitral annulus. Anterior leaflet coaptation length was measured under conditions of left ventricular saline distention: (1) at baseline, (2) with the sizer in place temporarily holding the sutures in position, and (3) after final implantation of a (corresponding size) Physio II ring (Edwards Lifesciences, Irivne, CA, USA). The “predicted” (sizer) and “actual” (implanted Physio II ring) coaptation lengths were compared using Bland–Altman analysis to assess accuracy.

Results

A moderate (2.0 to 5.5 mm) increase in coaptation length was achieved in 19 hearts and no change (0.0 mm) in 1 heart. The initial predicted (sizer) coaptation lengths corresponded well (≤0.5 mm difference) with the final, actual (implanted ring) coaptation lengths across all 20 hearts. Bland–Altman plots are provided for (derived) coaptation length increases. The mean difference (95% confidence interval) indicated that predicted were slightly greater than actual (0.08 [−0.06, 0.21]) but not different than zero. Limits of agreement were from −0.50 to 0.65. There did not appear to be significant heteroscedasticity.

Conclusions

The accuracy of a novel mitral annuloplasty ring-sizing device was validated using a cadaveric swine heart model. Coaptation lengths initially predicted by the sizer corresponded accurately with coaptation lengths observed upon implantation of a corresponding commercially available ring. Continued development of this technology should provide surgeons with precise control over the final coaptation length by previewing several different rings prior to committing to a final implant.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Mitral Valve AnnuloplastyMitral ValveHeart Valve ProsthesisAnimalsSwineModels, AnimalProsthesis DesignCadaver

Stichwörter der Autor:innen

annuloplastycoaptation lengthmitral repairpaired ringsizersizingtricuspid repair

Autor:innen und Institutionen

  • Division of Cardiac Surgery, Department of Surgery, NorthShore University Hospital System, Chicago, IL, USA.

ORCID-Profile: Nikola Dobrilovic

Zitieren

APA

Dobrilovic, N. (2026). Physiologic Mitral Annuloplasty Ring Sizing: Surgical Technique and Animal Model Validation. Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery, 21(2), 159-165. https://doi.org/10.1177/15569845261437227

Vancouver

Dobrilovic N. Physiologic Mitral Annuloplasty Ring Sizing: Surgical Technique and Animal Model Validation. Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery. 2026;21(2):159-165. doi:10.1177/15569845261437227

BibTeX
@article{Dobrilovic2026,
  author = {Dobrilovic, Nikola},
  title = {Physiologic Mitral Annuloplasty Ring Sizing: Surgical Technique and Animal Model Validation},
  journal = {Innovations: Technology and Techniques in Cardiothoracic and Vascular Surgery},
  year = {2026},
  volume = {21},
  number = {2},
  pages = {159--165},
  doi = {10.1177/15569845261437227}
}
Kostenlos starten →