Mixed Reality Surgical Navigation System; Positional Accuracy Based on Food and Drug Administration Standard
Surgical Innovation · 2024-02 · 31 (1) · 48-57
DOI: 10.1177/15533506231217620 ↗Kurz gefasst
Fachartikel in der Zeitschrift Surgical Innovation, online erschienen am 29.11.2023. Verfasst von Christopher T Morley und 5 weiteren Autor:innen. Beteiligte Einrichtung: Medivis, Inc., New York, NY, USA. Hauptthemen laut MEDLINE: Augmented Reality, Surgery, Computer-Assisted. Die Arbeit wurde bisher 22-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background
Computer assisted surgical navigation systems are designed to improve outcomes by providing clinicians with procedural guidance information. The use of new technologies, such as mixed reality, offers the potential for more intuitive, efficient, and accurate procedural guidance. The goal of this study is to assess the positional accuracy and consistency of a clinical mixed reality system that utilizes commercially available wireless head-mounted displays (HMDs), custom software, and localization instruments.
Methods
Independent teams using the second-generation Microsoft HoloLens © hardware, Medivis SurgicalAR © software, and localization instruments, tested the accuracy of the combined system at different institutions, times, and locations. The ASTM F2554-18 consensus standard for computer-assisted surgical systems, as recognized by the U.S. FDA, was utilized to measure the performance. 288 tests were performed.
Results
The system demonstrated consistent results, with an average accuracy performance that was better than one millimeter (.75 ± SD .37 mm).
Conclusion
Independently acquired positional tracking accuracies exceed conventional in-market surgical navigation tracking systems and FDA standards. Importantly, the performance was achieved at two different institutions, using an international testing standard, and with a system that included a commercially available off-the-shelf wireless head mounted display and software.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Medivis, Inc., New York, NY, USA.
- US Department of Veterans Affairs, Palo Alto Healthcare System, Palo Alto, CA, USA.
- Department of Business Analytics, Tippie College of Business, University of Iowa, Iowa, IA, USA.
- Department of Radiology, Stanford University School of Medicine, Stanford, CA, USA.
ORCID-Profile: David M Arreola, Thomas F Osborne
Zitieren
APA
Morley, C. T., Arreola, D. M., Qian, L., Lynn, A. L., Veigulis, Z. P., & Osborne, T. F. (2024). Mixed Reality Surgical Navigation System; Positional Accuracy Based on Food and Drug Administration Standard. Surgical Innovation, 31(1), 48-57. https://doi.org/10.1177/15533506231217620
Vancouver
Morley CT, Arreola DM, Qian L, Lynn AL, Veigulis ZP, Osborne TF. Mixed Reality Surgical Navigation System; Positional Accuracy Based on Food and Drug Administration Standard. Surgical Innovation. 2024;31(1):48-57. doi:10.1177/15533506231217620
BibTeX
@article{Morley2024,
author = {Morley, Christopher T and Arreola, David M and Qian, Long and Lynn, Amy L and Veigulis, Zachary P and Osborne, Thomas F},
title = {Mixed Reality Surgical Navigation System; Positional Accuracy Based on Food and Drug Administration Standard},
journal = {Surgical Innovation},
year = {2024},
volume = {31},
number = {1},
pages = {48--57},
doi = {10.1177/15533506231217620}
}