Use of the ExoSeal Vascular Closure Device in Patients With Peripheral Artery Disease: Risk Analysis of Deployment Failure and Prolonged Hemostasis

Ryo Okusako iD, Taira Kobayashi iD, Takanobu Okazaki, Masaki Hamamoto, Shinya Takahashi

Vascular and Endovascular Surgery · 2025

DOI: 10.1177/15385744251409971 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Vascular and Endovascular Surgery, 2025. Verfasst von Ryo Okusako und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Surgery, JA Hiroshima General Hospital, Hatsukaichi, Japan. Hauptthemen laut MEDLINE: Peripheral Arterial Disease, Hemostatic Techniques, Vascular Closure Devices, Endovascular Procedures. Die Arbeit nennt 14 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Endovascular treatment (EVT) of lower extremity artery disease (LEAD) is minimally invasive and has good safety and efficacy. Puncture site complications are common in EVT, but these and the hemostatic time have been reduced by newer hemostatic devices such as ExoSeal. However, factors affecting poor outcomes after use of ExoSeal are unclear. Therefore, the purpose of this study is to identify these factors after ExoSeal deployment in EVT. Material and Methods A retrospective study was performed for 868 cases in which hemostasis was attempted using ExoSeal between 2019 and 2023. Factors affecting deployment failure and prolonged hemostatic time after ExoSeal were assessed.

Results

The 868 subjects (569 males, 66%) had a median age of 77 [72-83] years. A femoral antegrade approach was used in 548 cases (63%), with use of a 6Fr sheath in most cases (99%). The lesions were in the femoropopliteal ( n = 387, 44%) and infrapopliteal ( n = 359, 41%) arteries. The technical success rate of EVT was 97%. ExoSeal was successfully deployed in 812 cases (94%). Cases in which ExoSeal deployment was unsuccessful had significantly more frequent proximal stenting (9% vs 20%, P = .01). In the entire cohort, there were 6 (.69%) puncture site complications. Of the 812 cases with successful deployment, the manual compression hemostasis time was prolonged (>10 min) in 51 cases (6%). Hemostatic time was significantly prolonged in cases with proximal stenting or intimal thickening at the puncture site (both, P < .001).

Conclusion

These results show that ExoSeal is a useful device. However, deployment failure and prolonged hemostatic time were caused by a proximal stent or intimal thickening of the puncture site, and use of hemostatic devices in such cases requires further investigation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Peripheral Arterial DiseaseHemostatic TechniquesVascular Closure DevicesEndovascular ProceduresHemorrhageHumansMaleRetrospective StudiesFemaleAgedRisk FactorsAged, 80 and overTime FactorsPuncturesRisk AssessmentTreatment OutcomeEquipment Design

Stichwörter der Autor:innen

ExoSealendovascular treatmentperipheral artery diseasevascular closure device

Autor:innen und Institutionen

  • Department of Cardiovascular Surgery, JA Hiroshima General Hospital, Hatsukaichi, Japan.
  • Department of Cardiovascular Surgery, Hiroshima University Hospital, Hiroshima, Japan.

ORCID-Profile: Ryo Okusako, Taira Kobayashi

Zitieren

APA

Okusako, R., Kobayashi, T., Okazaki, T., Hamamoto, M., & Takahashi, S. (2025). Use of the ExoSeal Vascular Closure Device in Patients With Peripheral Artery Disease: Risk Analysis of Deployment Failure and Prolonged Hemostasis. Vascular and Endovascular Surgery. https://doi.org/10.1177/15385744251409971

Vancouver

Okusako R, Kobayashi T, Okazaki T, Hamamoto M, Takahashi S. Use of the ExoSeal Vascular Closure Device in Patients With Peripheral Artery Disease: Risk Analysis of Deployment Failure and Prolonged Hemostasis. Vascular and Endovascular Surgery. 2025. doi:10.1177/15385744251409971

BibTeX
@article{Okusako2025,
  author = {Okusako, Ryo and Kobayashi, Taira and Okazaki, Takanobu and Hamamoto, Masaki and Takahashi, Shinya},
  title = {Use of the ExoSeal Vascular Closure Device in Patients With Peripheral Artery Disease: Risk Analysis of Deployment Failure and Prolonged Hemostasis},
  journal = {Vascular and Endovascular Surgery},
  year = {2025},
  doi = {10.1177/15385744251409971}
}
Kostenlos starten →