Early Results of the Newly-Designed Flowdynamics Dense Mesh Stent for Residual Dissection After Proximal Repair of Stanford Type A or Type B Aortic Dissection: A Preliminary Single-Center Report From a Multicenter, Prospective, and Randomized Study
Journal of Endovascular Therapy · 2024
DOI: 10.1177/15266028231163057 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Journal of Endovascular Therapy, 2024. Verfasst von Chen Lu und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Surgery, West China Hospital, Sichuan University, Chengdu, People's Republic of China. Hauptthemen laut MEDLINE: Aortic Dissection, Stents, Prosthesis Design, Endovascular Procedures. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 40 Literaturangaben. Gefördert von: sichuan province science and technology support program, National Natural Science Foundation of China, 1·3·5 project for disciplines of excellence, West China Hospital, Sichuan University.
Abstract (Originalsprache)
Background
Negative remodeling of the distal aorta due to residual dissection significantly impacts the long-term outcomes of dissection patients after proximal repair of acute aortic dissection. Branched/fenestrated aortic stents are technically demanding, and studies of the first generation of multilayer flow modulators for tackling this clinical scenario are few and limited. The single-center results from a multicenter, prospective, and randomized controlled study aimed to verify the safety and effectiveness of a newly-designed flowdynamics dense mesh stent for treating residual dissection after proximal repair.
Methods
Patients with nonchronic residual dissection involving visceral branches were prospectively enrolled in 3 centers (ChiCTR1900023638). Eligible patients were randomly assigned to the flowdynamics dense mesh stent (FDMS) group and control group. Follow-up visits were arranged at 1, 3, 6, and 12 months after recruitment. The primary endpoints were all-cause and aortic-related mortality. The secondary endpoints included visceral branch occlusion, reintervention, and severe adverse events. Morphological changes were analyzed to exhibit the therapeutic effect. Our center participated in the multicenter prospective randomized controlled trial, and the preliminary single-center experience was reported.
Results
Thirty-six patients were enrolled in our center, and the baseline characteristics of the 2 groups were comparable. Thirty-four patients completed the 12 month follow-up. Freedom from all-cause and aortic-related death were 94.4% and 100%. All visceral branches remained patent in the FDMS group. Increased area of the true lumen (1.03±0.38 vs 0.48±0.63 cm 2 at the plane below renal arteries, p=0.006; 1.27±0.80 vs 0.32±0.50 cm 2 at the plane 5 cm below renal arteries, p
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiovascular Surgery, West China Hospital, Sichuan University, Chengdu, People's Republic of China.
- Department of Cardiothoracic Surgery, Guang'an Hospital, West China Hospital, Sichuan University, Guang'an, People's Republic of China.
ORCID-Profile: Jia Hu
Förderung
- sichuan province science and technology support program
- National Natural Science Foundation of China
- 1·3·5 project for disciplines of excellence, West China Hospital, Sichuan University
- 1·3·5 project for disciplines of excellence–Clinical Research Incubation Project, West China Hospital, Sichuan University
Zitieren
APA
Lu, C., Wang, H., Yang, P., Liu, Y., Zhang, Y., Xu, Z., Xie, Y., & Hu, J. (2024). Early Results of the Newly-Designed Flowdynamics Dense Mesh Stent for Residual Dissection After Proximal Repair of Stanford Type A or Type B Aortic Dissection: A Preliminary Single-Center Report From a Multicenter, Prospective, and Randomized Study. Journal of Endovascular Therapy. https://doi.org/10.1177/15266028231163057
Vancouver
Lu C, Wang H, Yang P, Liu Y, Zhang Y, Xu Z, et al. Early Results of the Newly-Designed Flowdynamics Dense Mesh Stent for Residual Dissection After Proximal Repair of Stanford Type A or Type B Aortic Dissection: A Preliminary Single-Center Report From a Multicenter, Prospective, and Randomized Study. Journal of Endovascular Therapy. 2024. doi:10.1177/15266028231163057
BibTeX
@article{Lu2024,
author = {Lu, Chen and Wang, Haiyue and Yang, Peng and Liu, Yu and Zhang, Yu and Xu, Zhenyuan and Xie, Yi and Hu, Jia},
title = {Early Results of the Newly-Designed Flowdynamics Dense Mesh Stent for Residual Dissection After Proximal Repair of Stanford Type A or Type B Aortic Dissection: A Preliminary Single-Center Report From a Multicenter, Prospective, and Randomized Study},
journal = {Journal of Endovascular Therapy},
year = {2024},
doi = {10.1177/15266028231163057}
}
EUVIMED