Accuracy of a Continuous Glucose Monitor in the Intensive Care Unit
Genauigkeit eines kontinuierlichen Glukosemonitors auf der Intensivstation
Diabetes Technology & Therapeutics · 01.09.2026 · 28 (9) · 916-923
DOI: 10.1177/15209156261423933 ↗Auf einen Blick
Bei 16 Intensivpatient:innen mit Hyperglykämie zeigte der Dexcom G7 im Vergleich zu Routinewerten eine hohe Übereinstimmung. Die mittlere absolute relative Differenz betrug 12,25 %.
Fragestellung
Wie genau sind CGM-Glukosemessungen auf der Intensivstation im Vergleich zu Standardmessungen aus der Routineversorgung?
Methode
Nicht randomisierte Studie mit 16 informierten Intensivpatient:innen; Dexcom G7; Vergleich mit nahezu zeitgleichen Labor- oder Point-of-care-Messungen; Genauigkeit mit Clarke Error Grid und MARD beurteilt.
Ergebnisse
941 nahezu zeitgleiche Messpaare. 99,7 % der Werte lagen im Clarke Error Grid in den Regionen A oder B, 0,3 % in anderen Regionen. MARD 12,25 %. Keine dem CGM zuzuordnenden unerwünschten Ereignisse.
Einschränkungen
Nicht randomisiert; kleine Stichprobe; qualitative Akzeptanzbeurteilung.
Schlussfolgerung der Autor:innen
CGM-Glukosewerte auf der Intensivstation können in bestimmten kritisch kranken Patientengruppen zuverlässig sein. Bei Patient:innen mit Bedarf an hohen Vasopressordosen könnte CGM weniger zuverlässig sein.
Kernaussagen
- 941 Messpaare wurden verglichen
- 99,7 % im Clarke Error Grid in A oder B
- MARD 12,25 %; keine CGM-bedingten unerwünschten Ereignisse
Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.
Kurz gefasst
Fachartikel in der Zeitschrift Diabetes Technology & Therapeutics, online erschienen am 11.02.2026. Verfasst von Eric Cucchi und 5 weiteren Autor:innen. Beteiligte Einrichtung: UMass Chan Medical School, Worcester, Massachusetts, USA. Hauptthemen laut MEDLINE: Continuous Glucose Monitoring, Intensive Care Units, Blood Glucose, Hyperglycemia. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 26 Literaturangaben. Gefördert von: UMass Memorial Health DigiGrant.
Themen: Intensivpflege
Abstract (Originalsprache)
Objectives
Continuous glucose monitoring (CGM) devices are routinely used in the outpatient management of diabetes. They are not yet approved for use in intensive care units (ICUs). The main objective of this study was to determine the accuracy of CGM glucose determinations made in an ICU by comparing them with standard glucose measurements obtained during routine care. Secondary objectives were to determine the frequency of device malfunctions or adverse events and to assess acceptance of CGM use by caregivers.
Methods
Dexcom™ G7 CGM sensors were placed on non-randomized, consented ICU patients with known hyperglycemia. CGM glucose concentrations were recorded in a Dexcom Clarity™ database and compared with near-simultaneous (±5 min) measurements obtained by laboratory or point-of-care measurements recorded in an electronic medical record. Determinations of accuracy were made using a Clarke Error Grid plot and by calculating mean absolute relative difference (MARD). Device failure and adverse event data were recorded in the electronic record. Caregiver acceptance was assessed by interview.
Results
During a 9-month period, 16 subjects were enrolled after obtaining informed consent. The median duration of CGM use on subjects was 5.5 (range 1–20) days. A total of 941 near-simultaneous CGM and routine care data were collected. Clarke Error Grid comparison of CGM and routine care data showed 99.7% of readings in regions A or B and 0.3% in other regions. The MARD of all readings was 12.25%. There were no adverse events attributable to the CGM recorded. Qualitative analysis of nursing responses revealed no concerns regarding the addition of CGM to the routine workflow.
Conclusion
CGM glucose readings in the ICU may be reliable to use in certain critically ill patient populations. CGM may not be reliable in patients requiring high-dose vasopressors. The use of CGM to enhance patient care in ICUs merits additional research.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- UMass Chan Medical School, Worcester, Massachusetts, USA.
ORCID-Profile: Eric Cucchi
Förderung
- UMass Memorial Health DigiGrant
Zitieren
APA
Cucchi, E., Ali, A., Guilarte-Walker, Y., Barton, B., Malkani, S., & Mordes, J. (2026). Accuracy of a Continuous Glucose Monitor in the Intensive Care Unit. Diabetes Technology & Therapeutics, 28(9), 916-923. https://doi.org/10.1177/15209156261423933
Vancouver
Cucchi E, Ali A, Guilarte-Walker Y, Barton B, Malkani S, Mordes J. Accuracy of a Continuous Glucose Monitor in the Intensive Care Unit. Diabetes Technology & Therapeutics. 2026;28(9):916-923. doi:10.1177/15209156261423933
BibTeX
@article{Cucchi2026,
author = {Cucchi, Eric and Ali, Asem and Guilarte-Walker, Yurima and Barton, Bruce and Malkani, Samir and Mordes, John},
title = {Accuracy of a Continuous Glucose Monitor in the Intensive Care Unit},
journal = {Diabetes Technology & Therapeutics},
year = {2026},
volume = {28},
number = {9},
pages = {916--923},
doi = {10.1177/15209156261423933}
}
EUVIMED