Development of a Diagnostic Tool for Elbow Instability Combining 4D-CT Imaging and Mechanical Load Analysis

Jonas Gehring, Fabian Forster, Markus Meindl, Martin Maerz, Michael Simon, Nina Renner, Michael Uder, Tobias Pogarell, Lars Braeuer, Mario Perl, Lucas Mangold iD

Biomedical Engineering and Computational Biology · 2026-08 · 17

DOI: 10.1177/11795972261472224 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Engineering and Computational Biology, online erschienen am 10.08.2026. Verfasst von Jonas Gehring und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Power Electronics, Fraunhofer Institute for Integrated Systems and Device Technology IISB, Erlangen, Germany. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Siemens Healthineers.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Diagnosis of elbow instability in clinical practice commonly relies on manual stress testing by the physician with or without combination of fluoroscopy or ultrasound, which may result in variable force application and subjective interpretation. A standardized method for applying controlled mechanical loading during dynamic imaging may improve reproducibility.

Objective

The objective of this study was to develop and perform initial cadaveric feasibility testing of a CT-compatible diagnostic device for device-guided assessment of elbow instability with simultaneous force and positional measurement.

Design

Technical development study with initial cadaveric feasibility testing.

Methods

Building upon the initial prototype the present paper, describes the development of MELBO the follow-up prototype with adjusted kinematics and improved stability, the validation of the force sensor is addressed, and the first cadaver test results are presented.

Results

The MELBO diagnostic tool enabled a controlled device-guided elbow motion during 4D-CT acquisition and allowed simultaneous recording of force and positional data. The acquired dynamic CT datasets were suitable for qualitative assessment of elbow joint stability. No relevant device-related imaging artifacts were observed.

Conclusion

The present study demonstrates the technical feasibility of device-guided simultaneous force and positional measurement during 4D-CT using the MELBO diagnostic tool. These findings provide the basis for further cadaveric studies and future clinical investigations aimed at quantitative assessment of elbow instability under standardized loading conditions.

Schlagwörter

Stichwörter der Autor:innen

4D-CT imagingMelbodiagnostic toolelbow instabilitymechanical load analysis

Autor:innen und Institutionen

  • Department of Power Electronics, Fraunhofer Institute for Integrated Systems and Device Technology IISB, Erlangen, Germany.
  • Department of Power Electronics at Friedrich-Alexander-Universität Erlangen-Nürnberg, Nürnberg, Germany.
  • Department of Orthopaedics and Trauma Surgery, University Hospital of Erlangen, Friedrich-Alexander-University Erlangen-Nuremberg, Erlangen, Germany.
  • Department of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany.
  • Institute of Functional and Clinical Anatomy, Friedrich-Alexander-Universität (FAU) Erlangen-Nürnberg, Erlangen, Germany.

ORCID-Profile: Lucas Mangold

Förderung

  • Siemens Healthineers

Zitieren

APA

Gehring, J., Forster, F., Meindl, M., Maerz, M., Simon, M., Renner, N., Uder, M., Pogarell, T., Braeuer, L., Perl, M., & Mangold, L. (2026). Development of a Diagnostic Tool for Elbow Instability Combining 4D-CT Imaging and Mechanical Load Analysis. Biomedical Engineering and Computational Biology, 17. https://doi.org/10.1177/11795972261472224

Vancouver

Gehring J, Forster F, Meindl M, Maerz M, Simon M, Renner N, et al. Development of a Diagnostic Tool for Elbow Instability Combining 4D-CT Imaging and Mechanical Load Analysis. Biomedical Engineering and Computational Biology. 2026;17. doi:10.1177/11795972261472224

BibTeX
@article{Gehring2026,
  author = {Gehring, Jonas and Forster, Fabian and Meindl, Markus and Maerz, Martin and Simon, Michael and Renner, Nina and Uder, Michael and Pogarell, Tobias and Braeuer, Lars and Perl, Mario and Mangold, Lucas},
  title = {Development of a Diagnostic Tool for Elbow Instability Combining 4D-CT Imaging and Mechanical Load Analysis},
  journal = {Biomedical Engineering and Computational Biology},
  year = {2026},
  volume = {17},
  doi = {10.1177/11795972261472224}
}
Kostenlos starten →