Numerical and Clinical Accuracy of the FreeStyle Libre 2 System: A Diagnostic Accuracy Study in Critically Ill Patients

Miguel Aguilar-Schotborgh iD, Oscar Muñoz-Velandia iD, Diana Cristina Henao-Carrillo iD, Luis Triana Moreno iD, Jhon Moreno Mogollon iD, Ana Maria Gómez iD

Clinical Medicine Insights: Endocrinology and Diabetes · 2026-07 · 19

DOI: 10.1177/11795514261470879 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Medicine Insights: Endocrinology and Diabetes, online erschienen am 18.07.2026. Verfasst von Miguel Aguilar-Schotborgh und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Internal Medicine, Pontificia Universidad Javeriana, Bogotá, Colombia. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The accuracy of the FreeStyle Libre 2 system in critically ill patients needing insulin therapy remains inadequately evaluated.

Objective

To evaluate the clinical and numerical accuracy of continuous glucose monitoring (FreeStyle Libre 2) and capillary glucometry (StatStrip) compared with central laboratory glucose in critically ill patients requiring insulin therapy.

Methods

We conducted a diagnostic accuracy study, evaluating simultaneously two index tests, the FreeStyle Libre 2 and capillary glucose, compared to the central laboratory glucose (venous or arterial) as reference standard. The study included critically ill adult patients admitted to the intensive care unit (ICU) with diabetes or stress hyperglycemia, requiring insulin therapy, and ventilatory or vasopressor support. Numerical accuracy was assessed using ISO 15197:2013 criteria and Mean Absolute Relative Difference (MARD). Clinical accuracy was evaluated using Clarke and Parkes error grid analysis.

Results

A total of 157 paired measurements were collected. FreeStyle Libre 2 showed a MARD comparable to capillary glucose (10.43% vs 7.58%; p=0.14). Neither method met ISO 15197:2013 numerical accuracy criteria. In the clinical accuracy analysis, FreeStyle Libre 2 classified 100% of measurements within zones A+B on both the Clarke and Parkes grids, whereas capillary glucose achieved 97.6% and 100%, respectively.

Conclusions

FreeStyle Libre 2 showed numerical accuracy comparable to capillary blood glucose and consistently reliable clinical performance in critically ill patients needing insulin therapy. These results support its potential as an alternative to capillary glucose monitoring in the ICU. More studies involving a greater number of hypoglycaemic events are required to confirm its effectiveness in this critical range.

Schlagwörter

Stichwörter der Autor:innen

ICUcontinuous glucose monitoringcritical carediabetes mellitusglycemic control

Autor:innen und Institutionen

  • Department of Internal Medicine, Pontificia Universidad Javeriana, Bogotá, Colombia.
  • Department of Internal Medicine, Hospital Universitario San Ignacio, Bogotá, Colombia.
  • Endocrinology Unit, Hospital Universitario San Ignacio, Bogotá, Colombia.
  • Critical Care Unit, Hospital Universitario San Ignacio, Bogotá, Colombia.

ORCID-Profile: Miguel Aguilar-Schotborgh, Oscar Muñoz-Velandia, Diana Cristina Henao-Carrillo, Luis Triana Moreno, Jhon Moreno Mogollon, Ana Maria Gómez

Zitieren

APA

Aguilar-Schotborgh, M., Muñoz-Velandia, O., Henao-Carrillo, D. C., Triana Moreno, L., Moreno Mogollon, J., & Gómez, A. M. (2026). Numerical and Clinical Accuracy of the FreeStyle Libre 2 System: A Diagnostic Accuracy Study in Critically Ill Patients. Clinical Medicine Insights: Endocrinology and Diabetes, 19. https://doi.org/10.1177/11795514261470879

Vancouver

Aguilar-Schotborgh M, Muñoz-Velandia O, Henao-Carrillo DC, Triana Moreno L, Moreno Mogollon J, Gómez AM. Numerical and Clinical Accuracy of the FreeStyle Libre 2 System: A Diagnostic Accuracy Study in Critically Ill Patients. Clinical Medicine Insights: Endocrinology and Diabetes. 2026;19. doi:10.1177/11795514261470879

BibTeX
@article{AguilarSchotborgh2026,
  author = {Aguilar-Schotborgh, Miguel and Muñoz-Velandia, Oscar and Henao-Carrillo, Diana Cristina and Triana Moreno, Luis and Moreno Mogollon, Jhon and Gómez, Ana Maria},
  title = {Numerical and Clinical Accuracy of the FreeStyle Libre 2 System: A Diagnostic Accuracy Study in Critically Ill Patients},
  journal = {Clinical Medicine Insights: Endocrinology and Diabetes},
  year = {2026},
  volume = {19},
  doi = {10.1177/11795514261470879}
}
Kostenlos starten →