Retrospective study of vacuum aspiration mechanical thrombectomy device in the treatment of arteriovenous fistula thrombosis and patient is on hemodialysis
The Journal of Vascular Access · 2026
DOI: 10.1177/11297298251326979 ↗Kurz gefasst
Fachartikel in der Zeitschrift The Journal of Vascular Access, 2026. Verfasst von Glauber Lutterbach de Oliveira Pires und 6 weiteren Autor:innen. Beteiligte Einrichtung: Programa de Pós-Graduação em Cirurgia/CCM-UFPE, Recife, Pernambuco, Brazil. Hauptthemen laut MEDLINE: Renal Dialysis, Arteriovenous Shunt, Surgical, Thrombectomy, Graft Occlusion, Vascular. Die Arbeit nennt 30 Literaturangaben.
Abstract (Originalsprache)
Objective
To evaluate the use of the device of mechanical thrombectomy aspiration in the salvage of the fistula for hemodialysis access (HD) with acute thrombosis.
Method
The study model used was a retrospective cohort in which 35 patients, who underwent 44 percutaneus thrombectomy in a tertiary hospital between October 2018 and March 2020, were evaluated. The surgical time of the thrombectomy, the technical success (TS) and the clinical success (CS) were evaluated, as well as the occurrence of surgical complications. Additional adjunct therapy, such balloon angioplasty or thrombolytic therapy when performed was not evaluated.
Results
Among the 35 evaluated patients, 24 (66.8%) were male. In 33 of these cases (75%), the fistula used an autologous vein (AV fistula). The average thrombectomy time was 20.2 min and median was 15 min, with a difference between the AV fistula and synthetic graft (Graft) ( p = 0.001), being shorter in the Graft. The TS rate was of 86.4% and the CS rate was of 84.1% with no statistical difference between the groups. The rate of surgical complications was of 13.6%. Blood loss in most cases ranged from 100 to 300 ml in the majority of the cases.
Conclusion
The Device showed a short operation time in its use, a high rate of technical success and clinical success, as well as a similar rate of immediate surgical complications found in the literature.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Programa de Pós-Graduação em Cirurgia/CCM-UFPE, Recife, Pernambuco, Brazil.
- Programa de Pós-graduação em Recursos Naturais do Semiárido, UNIFASV, Petrolina, Pernambuco, Brazil.
- Vascular Surgeon and Interventional Radiologist, Angiorad, Recife, Pernambuco, Brazil.
- Área Acadêmica de Cirurgia, CCM-UFPE, Recife, Pernambuco, Brazil.
ORCID-Profile: Esdras Marques Lins
Zitieren
APA
Pires, G. L. O., Moreira Lutterbach, A. C. T., Rivera, M. A. C. P., Cavalcanti, D., Ayub, J. P., Appolonio, F., & Lins, E. M. (2026). Retrospective study of vacuum aspiration mechanical thrombectomy device in the treatment of arteriovenous fistula thrombosis and patient is on hemodialysis. The Journal of Vascular Access. https://doi.org/10.1177/11297298251326979
Vancouver
Pires GLO, Moreira Lutterbach ACT, Rivera MACP, Cavalcanti D, Ayub JP, Appolonio F, et al. Retrospective study of vacuum aspiration mechanical thrombectomy device in the treatment of arteriovenous fistula thrombosis and patient is on hemodialysis. The Journal of Vascular Access. 2026. doi:10.1177/11297298251326979
BibTeX
@article{Pires2026,
author = {Pires, Glauber Lutterbach de Oliveira and Moreira Lutterbach, Ana Claudia Teixeira and Rivera, Marco Antônio Cassiano Perez and Cavalcanti, Douglas and Ayub, João Paulo and Appolonio, Fernanda and Lins, Esdras Marques},
title = {Retrospective study of vacuum aspiration mechanical thrombectomy device in the treatment of arteriovenous fistula thrombosis and patient is on hemodialysis},
journal = {The Journal of Vascular Access},
year = {2026},
doi = {10.1177/11297298251326979}
}
EUVIMED