EUVIMEDEUVIMED entdecken
Beobachtungsstudie

Anticipating impending peripheral intravenous catheter failure: A diagnostic accuracy observational study combining ultrasound and artificial intelligence to improve clinical care

Amit Bahl iD, Steven Johnson, Nicholas Mielke, Michael Blaivas, Laura Blaivas

The Journal of Vascular Access · 2025

DOI: 10.1177/11297298241307055 ↗

Kurz gefasst

Beobachtungsstudie in der Zeitschrift The Journal of Vascular Access, 2025. Verfasst von Amit Bahl und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Emergency Medicine, Beaumont Hospital, Royal Oak, MI, USA. Hauptthemen laut MEDLINE: Catheterization, Peripheral, Machine Learning, Catheters, Indwelling, Artificial Intelligence. Die Arbeit nennt 19 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

Peripheral intravenous catheter (PIVC) failure occurs in approximately 50% of insertions. Unexpected PIVC failure leads to treatment delays, longer hospitalizations, and increased risk of patient harm. In current practice there is no method to predict if PIVC failure will occur until it is too late and a grossly obvious complication has occurred. The aim of this study is to demonstrate the diagnostic accuracy of a predictive model for PIVC failure based on artificial intelligence (AI).

Methods

This study evaluated the capabilities of a novel machine learning algorithm. The algorithm was trained using real-world ultrasound videos of PIVC sites with a goal of predicting which PIVCs would fail within the following day. After training, AI models were validated using another, unseen, collection of real-world ultrasound videos of PIVC sites.

Results

2133 ultrasound videos (361 failure and 1772 non-failure) were used for algorithm development. When the algorithm was tasked with predicting failure in the unseen collection of videos, the best achieved results were an accuracy of 0.93, sensitivity of 0.77, specificity of 0.98, positive predictive value of 0.91, negative predictive value of 0.93, and area under the curve of 0.87.

Conclusions

This proprietary and novel machine learning algorithm can accurately and reliably predict PIVC failure 1 day prior to clinically evident failure. Implementation of this technology in the patient care setting would provide timely information for clinicians to plan and manage impending device failure. Future research on the use of AI technology and PIVCs should focus on improving catheter function and longevity, while limiting complication rates.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Catheterization, PeripheralMachine LearningCatheters, IndwellingArtificial IntelligenceVascular Access DevicesDecision Support TechniquesImage Interpretation, Computer-AssistedEquipment FailureQuality ImprovementHumansPredictive Value of TestsVideo RecordingReproducibility of ResultsRisk FactorsAlgorithmsUltrasonographyTime FactorsRisk Assessment

Stichwörter der Autor:innen

Artificial intelligenceassessmentcare and maintenanceduplexmachine learningperipheral intravenous catheter failureultrasound

Autor:innen und Institutionen

  • Department of Emergency Medicine, Beaumont Hospital, Royal Oak, MI, USA.
  • Department of Anesthesia Critical Care, University of Southern California, Los Angeles, CA, USA.
  • Department of Medicine, Creighton University School of Medicine, Omaha, NE, USA.
  • Department of Medicine, University of South Carolina School of Medicine, Columbia, SC, USA.
  • Department of Environmental Sciences, Michigan State University, Lansing, MI, USA.

ORCID-Profile: Amit Bahl

Zitieren

APA

Bahl, A., Johnson, S., Mielke, N., Blaivas, M., & Blaivas, L. (2025). Anticipating impending peripheral intravenous catheter failure: A diagnostic accuracy observational study combining ultrasound and artificial intelligence to improve clinical care. The Journal of Vascular Access. https://doi.org/10.1177/11297298241307055

Vancouver

Bahl A, Johnson S, Mielke N, Blaivas M, Blaivas L. Anticipating impending peripheral intravenous catheter failure: A diagnostic accuracy observational study combining ultrasound and artificial intelligence to improve clinical care. The Journal of Vascular Access. 2025. doi:10.1177/11297298241307055

BibTeX
@article{Bahl2025,
  author = {Bahl, Amit and Johnson, Steven and Mielke, Nicholas and Blaivas, Michael and Blaivas, Laura},
  title = {Anticipating impending peripheral intravenous catheter failure: A diagnostic accuracy observational study combining ultrasound and artificial intelligence to improve clinical care},
  journal = {The Journal of Vascular Access},
  year = {2025},
  doi = {10.1177/11297298241307055}
}
Kostenlos starten →