A 4-year tunneled hemodialysis catheter malpositioned into the azygos vein and how to remedy the hemodialysis circuit

Yuqin Xiong, Qiuyan Zhao, Yang Yu iD, Tianlei Cui iD

The Journal of Vascular Access · 2025

DOI: 10.1177/11297298231223108 ↗

Kurz gefasst

Fallbericht in der Zeitschrift The Journal of Vascular Access, 2025. Verfasst von Yuqin Xiong und 3 weiteren Autor:innen. Beteiligte Einrichtung: Kidney Research Institute, West China Hospital of Sichuan University, Chengdu, China. Hauptthemen laut MEDLINE: Angioplasty, Balloon, Azygos Vein, Catheterization, Central Venous, Catheters, Indwelling. Die Arbeit nennt 9 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Mispositioning in the azygos vein is a rare but hazardous complication of central venous catheterization. A patient was admitted for a dysfunctional hemodialysis tunneled cuffed catheter (TCC) placed in the azygos vein for 4 years. Computed tomography angiography revealed multiple sites of occlusion, including the superior vena cava (SVC), right and left innominate veins (IVs), and right femoral vein. Percutaneous transluminal angioplasty and a TCC replacement based on a segment-by-segment recanalizing strategy were performed. First, an 8-Fr sheath was inserted through the left femoral vein approach to retrogradely traverse the occlusive SVC followed by a guidewire extending to the occlusive left IV. A left transjugular 15-cm snare was inserted to capture the transfemoral guidewire and achieve recanalization from the left IV to the SVC. Second, a transjugular guidewire was advanced through the dysfunctional TCC yet shunted into the left IV due to the inability to cross the SVC. A left transfemoral 15-cm snare was inserted to capture the guidewire and achieve complete recanalization from the right internal jugular vein to the SVC. Balloons were passed over the guidewires to dilate the obstructive lesions sequentially, and a new TCC was inserted successfully with the tip positioned in the right atrium.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Angioplasty, BalloonAzygos VeinCatheterization, Central VenousCatheters, IndwellingCentral Venous CathetersRenal DialysisHumansComputed Tomography AngiographyDevice RemovalPhlebographyTime FactorsTreatment Outcome

Stichwörter der Autor:innen

Azygos veincatheter mispositionhemodialysis accessrecanalization

Autor:innen und Institutionen

  • Kidney Research Institute, West China Hospital of Sichuan University, Chengdu, China.
  • Department of Nephrology, West China School of Public Health and West China Fourth Hospital, Sichuan University, Chengdu, China.
  • General Practice Ward/International Medical Center Ward, General Practice Medical Center, West China Hospital, Sichuan University, Chengdu, China.
  • West China School of Nursing, Sichuan University, Chengdu, China.
  • Department of Nephrology, Kidney Research Institute, West China Hospital of Sichuan University, Chengdu, China.

ORCID-Profile: Yang Yu, Tianlei Cui

Zitieren

APA

Xiong, Y., Zhao, Q., Yu, Y., & Cui, T. (2025). A 4-year tunneled hemodialysis catheter malpositioned into the azygos vein and how to remedy the hemodialysis circuit. The Journal of Vascular Access. https://doi.org/10.1177/11297298231223108

Vancouver

Xiong Y, Zhao Q, Yu Y, Cui T. A 4-year tunneled hemodialysis catheter malpositioned into the azygos vein and how to remedy the hemodialysis circuit. The Journal of Vascular Access. 2025. doi:10.1177/11297298231223108

BibTeX
@article{Xiong2025,
  author = {Xiong, Yuqin and Zhao, Qiuyan and Yu, Yang and Cui, Tianlei},
  title = {A 4-year tunneled hemodialysis catheter malpositioned into the azygos vein and how to remedy the hemodialysis circuit},
  journal = {The Journal of Vascular Access},
  year = {2025},
  doi = {10.1177/11297298231223108}
}
Kostenlos starten →