Axillary vein catheterization using ultrasound guidance: A prospective randomized cross-over controlled simulation comparing standard ultrasound and new needle-pilot device
The Journal of Vascular Access · 2023-09 · 24 (5) · 1042-1050
DOI: 10.1177/11297298211063705 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift The Journal of Vascular Access, online erschienen am 30.12.2021. Verfasst von Nicolas Boulet und 6 weiteren Autor:innen. Beteiligte Einrichtung: Intensive Care Unit, Department of Anesthesiology, Critical Care, Pain and Emergency Medicine, Nîmes University Hospital, Nîmes, France. Hauptthemen laut MEDLINE: Axillary Vein, Catheterization, Central Venous. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 32 Literaturangaben.
Abstract (Originalsprache)
Background
Real-time ultrasound (US) guidance facilitates central venous catheterization in intensive care unit (ICU). New magnetic needle-pilot devices could improve efficiency and safety of central venous catheterization. This simulation trial was aimed at comparing venipuncture with a new needle-pilot device to conventional US technique.
Methods
In a prospective, randomized, simulation trial, 51 ICU physicians and residents cannulated the right axillary vein of a human torso mannequin with standard US guidance and with a needle-pilot system, in a randomized order. The primary outcome was the time from skin puncture to successful venous cannulation. The secondary outcomes were the number of skin punctures, the number of posterior wall puncture of the axillary vein, the number of arterial punctures, the number of needle redirections, the failure rate, and the operator comfort.
Results
Time to successful cannulation was shorter with needle-pilot US-guided technique (22 s (interquartile range (IQR) = 16–42) vs 25 s (IQR = 19–128); median of difference (MOD) = −9 s (95%-confidence interval (CI) −5, −22), p < 0.001). The rates of skin punctures, posterior wall puncture of axillary vein, and needle redirections were also lower ( p < 0.01). Comfort was higher in needle-pilot US-guided group on a 11-points numeric scale (8 (IQR = 8–9) vs 6 (IQR = 6–8), p < 0.001).
Conclusions
In a simulation model, US-guided axillary vein catheterization with a needle-pilot device was associated with a shorter time of successful cannulation and a decrease in numbers of skin punctures and complications. The results plea for investigating clinical performance of this new device.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Intensive Care Unit, Department of Anesthesiology, Critical Care, Pain and Emergency Medicine, Nîmes University Hospital, Nîmes, France.
- EA 2992 IMAGINE, Department of Anesthesiology, Critical Care, Pain and Emergency Medicine, Nîmes University Hospital, Nîmes, France.
- Department of Biostatistics-Bioinformatic, Hospices Civils de Lyon, Lyon, France.
ORCID-Profile: Nicolas Boulet
Zitieren
APA
Boulet, N., Bobbia, X., Gavoille, A., Louart, B., Lefrant, J. Y., Roger, C., & Muller, L. (2023). Axillary vein catheterization using ultrasound guidance: A prospective randomized cross-over controlled simulation comparing standard ultrasound and new needle-pilot device. The Journal of Vascular Access, 24(5), 1042-1050. https://doi.org/10.1177/11297298211063705
Vancouver
Boulet N, Bobbia X, Gavoille A, Louart B, Lefrant JY, Roger C, et al. Axillary vein catheterization using ultrasound guidance: A prospective randomized cross-over controlled simulation comparing standard ultrasound and new needle-pilot device. The Journal of Vascular Access. 2023;24(5):1042-1050. doi:10.1177/11297298211063705
BibTeX
@article{Boulet2023,
author = {Boulet, Nicolas and Bobbia, Xavier and Gavoille, Antoine and Louart, Benjamin and Lefrant, Jean Yves and Roger, Claire and Muller, Laurent},
title = {Axillary vein catheterization using ultrasound guidance: A prospective randomized cross-over controlled simulation comparing standard ultrasound and new needle-pilot device},
journal = {The Journal of Vascular Access},
year = {2023},
volume = {24},
number = {5},
pages = {1042--1050},
doi = {10.1177/11297298211063705}
}
EUVIMED