Patterns of wear, fracture, and re-revision in ceramic-on-ceramic total hip arthroplasty: a 31-year retrieval study
HIP International · 22.11.2026
DOI: 10.1177/11207000261489753 ↗Kurz gefasst
Fachartikel in der Zeitschrift HIP International, online erschienen am 22.11.2026. Verfasst von William G Blakeney und 4 weiteren Autor:innen. Die Arbeit nennt 33 Literaturangaben.
Abstract (Originalsprache)
Introduction
Ceramic-on-ceramic (CoC) total hip arthroplasty (THA) is commonly used in younger and more active patients due to its low wear characteristics. However, ceramic fracture and wear-related complications remain important causes of revision. This retrieval study aimed to characterise failure mechanisms and associated patient and implant factors in revised CoC bearings.
Methods
A retrospective retrieval analysis of alumina and zirconia-toughened alumina (ZTA) CoC THA bearings revised between 1989 and 2021 was performed. Retrievals were classified as non-fractured primary revisions, fractured primary revisions, or re-revisions. Macroscopic and microscopic assessment of fracture morphology and wear patterns was undertaken. Volumetric wear and radiographic component positioning were assessed in a subset. Associations between ceramic fracture and patient activity, implant characteristics, and head size were analysed using logistic regression.
Results
The most common indications for revision were aseptic loosening (35.5%), pain (34.1%), ceramic fracture (17.7%), infection (13.2%), and instability (11.8%). Ceramic fracture predominantly involved the acetabular liner (34 liner fractures versus 12 femoral head fractures). Higher patient activity level was associated with increased odds of fracture (moderate activity: OR 11.8; 95% CI, 2.66–52.7; high activity: OR 5.6; 95% CI, 1.24–25.0). ZTA femoral heads (OR 0.30; 95% CI,0.13–0.66) and larger head sizes were associated with lower odds of fracture ( p < 0.001). Differences in volumetric wear between materials were not significant after adjustment for time in situ . Re-revision cases demonstrated more extensive third-body wear and shorter implant survival.
Conclusions
Ceramic fracture in CoC THA most commonly involves the acetabular liner and is associated with patient activity and implant factors. Re-revision cases demonstrate substantial third-body wear and reduced implant survival.
Zitieren
APA
Blakeney, W. G., Jeyaraj, R. S., Kop, A. M., Beaver, R. J., & Pabbruwe, M. (2026). Patterns of wear, fracture, and re-revision in ceramic-on-ceramic total hip arthroplasty: a 31-year retrieval study. HIP International. https://doi.org/10.1177/11207000261489753
Vancouver
Blakeney WG, Jeyaraj RS, Kop AM, Beaver RJ, Pabbruwe M. Patterns of wear, fracture, and re-revision in ceramic-on-ceramic total hip arthroplasty: a 31-year retrieval study. HIP International. 2026. doi:10.1177/11207000261489753
BibTeX
@article{Blakeney2026,
author = {Blakeney, William G and Jeyaraj, Reuben S and Kop, Alan M and Beaver, Richard J and Pabbruwe, Moreica},
title = {Patterns of wear, fracture, and re-revision in ceramic-on-ceramic total hip arthroplasty: a 31-year retrieval study},
journal = {HIP International},
year = {2026},
doi = {10.1177/11207000261489753}
}
EUVIMED