Can’t stop the slide: factors associated with lag screw slide following cephalomedullary nail fixation of intertrochanteric hip fractures
HIP International · 2026-07 · 36 (4) · 652-660
DOI: 10.1177/11207000261426459 ↗Kurz gefasst
Fachartikel in der Zeitschrift HIP International, online erschienen am 04.04.2026. Verfasst von Christopher J Pettit und 8 weiteren Autor:innen. Beteiligte Einrichtung: NYU Langone Orthopedic Hospital, New York, NY, USA. Hauptthemen laut MEDLINE: Bone Screws, Bone Nails, Fracture Fixation, Intramedullary, Hip Fractures. Die Arbeit nennt 43 Literaturangaben.
Abstract (Originalsprache)
Objective
To examine factors associated with lag screw slide following fixation of intertrochanteric hip fractures with 1 type of cephalomedullary nail.
Methods
Retrospective review of patients operatively treated for intertrochanteric hip fractures (OTA/AO 31A1 and 31A2) with a single cephalomedullary nail (CMN) at a single academic medical centre between November 2014 and November 2023. CMN lag screw was placed in “dynamic” mode to allow for controlled collapse, or screw “slide.” Screw slide was defined as the difference in lateral prominence of the lag screw at latest follow up compared to its initial position. Patients were grouped based on the amount of screw slide (15 mm) and correlation analysis was performed.
Results
614 intertrochanteric hip fracture patients were identified (mean age 80.76 years; 72.3% female) with mean 6.2 months follow-up. Mean amount of slide was 3.77 ± 4.79 mm. 66.3% of patients had 15 mm. Univariate analysis demonstrated that slide >15 mm was associated with increased patient BMI ( p = 15 mm of slide were only taking vitamin D and calcium whereas those with 15 mm screw slide.
Conclusions
Excessive lag screw slide (>15 mm) was associated with higher patient BMI. Patients with higher BMIs should be monitored to identify excessive slide. Surgeons should attempt to keep the lag screw as close to the lateral cortex as possible. While the use of anti-osteoporotic therapy was associated with more slide, this was almost exclusively seen in patients only prescribed vitamin D and calcium.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- NYU Langone Orthopedic Hospital, New York, NY, USA.
- Jamaica Hospital Medical Center, New York, NY, USA.
ORCID-Profile: Sanjit Konda, Kenneth A Egol
Zitieren
APA
Pettit, C. J., Herbosa, C., Fisher, N. D., Ganta, A., Rivero, S., Tejwani, N. C., Leucht, P., Konda, S., & Egol, K. A. (2026). Can’t stop the slide: factors associated with lag screw slide following cephalomedullary nail fixation of intertrochanteric hip fractures. HIP International, 36(4), 652-660. https://doi.org/10.1177/11207000261426459
Vancouver
Pettit CJ, Herbosa C, Fisher ND, Ganta A, Rivero S, Tejwani NC, et al. Can’t stop the slide: factors associated with lag screw slide following cephalomedullary nail fixation of intertrochanteric hip fractures. HIP International. 2026;36(4):652-660. doi:10.1177/11207000261426459
BibTeX
@article{Pettit2026,
author = {Pettit, Christopher J and Herbosa, Carolyn and Fisher, Nina D and Ganta, Abhishek and Rivero, Steven and Tejwani, Nirmal C and Leucht, Philipp and Konda, Sanjit and Egol, Kenneth A},
title = {Can’t stop the slide: factors associated with lag screw slide following cephalomedullary nail fixation of intertrochanteric hip fractures},
journal = {HIP International},
year = {2026},
volume = {36},
number = {4},
pages = {652--660},
doi = {10.1177/11207000261426459}
}
EUVIMED