Does the French Paradox cementing technique improve or compromise femoral stem stability? An in vitro cadaver study
HIP International · 2025
DOI: 10.1177/11207000251379369 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift HIP International, 2025. Verfasst von Stefan A St George und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, London Health Sciences Centre University Hospital, London, ON, Canada. Hauptthemen laut MEDLINE: Cementation, Hip Prosthesis, Bone Cements, Arthroplasty, Replacement, Hip. Die Arbeit nennt 27 Literaturangaben. Gefördert von: Natural Sciences and Engineering Research Council of Canada, Canada Foundation for Innovation.
Abstract (Originalsprache)
Background
The French Paradox (FP) technique, characterised by line-to-line cementing, has recently been applied to polished tapered stems with favourable results despite the thin cement mantle created. However, it contravenes the minimum 2-mm cement mantle described in the standard (STD) cementing technique.
Aim
This experimental study aimed to compare the FP and STD cementing techniques by comparing the cement mantle thickness, micromotion, and alignment of a polished tapered stem under clinically relevant loads.
Methods
Stems were implanted into 4 pairs of human cadaveric hips. Femurs within each pair were randomly allocated to receive either the STD or FP technique. 5 linear variable displacement transducers (LVDTs) were used to measure the real-time motion of the stem relative to the femur. The specimens were mounted onto a joint motion simulator and cyclically loaded with axial compression (0–1,600 N) and internal torsion (0–15 Nm).
Results
No significant differences were found in permanent migration, initial and long-term inducible motion, or stem alignment between techniques. The FP technique resulted in a thinner cement mantle (∆1.0 ± 0.2 mm). The mean permanent migration was below 20 μm, and inducible motions were below 100 μm for both techniques.
Conclusions
Both FP and STD cementing techniques resulted in well-fixed stems with low migration and inducible motion.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Orthopaedic Surgery, London Health Sciences Centre University Hospital, London, ON, Canada.
- School of Mechatronic Engineering and Automation, Shanghai University, Shanghai, China.
- Department of Mechanical and Materials Engineering, Western University, London, ON, Canada.
- School of Biomedical Engineering, Western University, London, ON, Canada.
ORCID-Profile: Stefan A St George, Yizhao Li
Förderung
- Natural Sciences and Engineering Research Council of Canada
- Canada Foundation for Innovation
Zitieren
APA
St George, S. A., Li, Y., Hoffer, A., Vasarhelyi, E., Willing, R., & Lanting, B. (2025). Does the French Paradox cementing technique improve or compromise femoral stem stability? An in vitro cadaver study. HIP International. https://doi.org/10.1177/11207000251379369
Vancouver
St George SA, Li Y, Hoffer A, Vasarhelyi E, Willing R, Lanting B. Does the French Paradox cementing technique improve or compromise femoral stem stability? An in vitro cadaver study. HIP International. 2025. doi:10.1177/11207000251379369
BibTeX
@article{StGeorge2025,
author = {St George, Stefan A and Li, Yizhao and Hoffer, Alexander and Vasarhelyi, Edward and Willing, Ryan and Lanting, Brent},
title = {Does the French Paradox cementing technique improve or compromise femoral stem stability? An in vitro cadaver study},
journal = {HIP International},
year = {2025},
doi = {10.1177/11207000251379369}
}
EUVIMED