Vergleichsstudie

Robotic reaming, multiple screw fixation results in significantly improved initial acetabular cup stability

Enrico M Forlenza, Robert A Burnett, Alejandro A Espinoza Orías iD, Vasili Karas

HIP International · 2026

DOI: 10.1177/11207000251374145 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift HIP International, 2026. Verfasst von Enrico M Forlenza und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedic Surgery, Rush University Medical Center, Chicago, IL, USA. Hauptthemen laut MEDLINE: Bone Screws, Acetabulum, Robotic Surgical Procedures, Arthroplasty, Replacement, Hip. Die Arbeit nennt 29 Literaturangaben. Gefördert von: Stryker.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

This study aimed to compare the biomechanical performance of cups implanted with manual versus robotic reaming, single versus sequential reaming, as well as no screws, 1 screw and 2 screws.

Methods

A total of 48, 56-mm diameter 3D-printed porous acetabular cups were impacted into separate polyurethane foam bone models reamed with either manual or robotic arm reaming, and either single or sequential reaming techniques. Simultaneous torsion-compression loading yielded torque at failure, angular displacement at failure and torsional stiffness of acetabular cups fixed with either 0, 1 or 2 acetabular screws.

Results

Overall, torsional stiffness for robotically-reamed cups (both sequential and single) was 45% higher than those reamed via manual instrumentation (3609.0 vs. 2484.6 N·m/rad; p < 0.0001). With increasing number of acetabular screws, the torque at failure (zero screws: 60.1 N·m vs. 1 screw 67.0 N·m vs. 2 screws 97.9 N·m; p < 0.001) and angular displacement at failure (0 screws: 2.0° vs. 1 screw 1.7° vs. 2 screws 5.3°; p < 0.001) increased. There was no difference in the torque, angular displacement at failure or torsional stiffness observed between cups reamed via single or sequential reaming (all p > 0.05). Neither torque nor angular displacement showed differences at failure (all p > 0.05).

Conclusions

Robotically-reamed cups demonstrated nearly 50% greater torsional stiffness than those reamed manually, suggesting that robotic reaming results in fewer deviations in sphericity, thereby optimising the congruency of the acetabulum-cup interface.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Bone ScrewsAcetabulumRobotic Surgical ProceduresArthroplasty, Replacement, HipHip ProsthesisHumansBiomechanical PhenomenaProsthesis DesignMaterials Testing

Stichwörter der Autor:innen

Acetabular componenthip arthroplastyreamingrobotic surgerytorsional testing

Autor:innen und Institutionen

  • Department of Orthopedic Surgery, Rush University Medical Center, Chicago, IL, USA.

ORCID-Profile: Alejandro A Espinoza Orías

Förderung

  • Stryker

Zitieren

APA

Forlenza, E. M., Burnett, R. A., Espinoza Orías, A. A., & Karas, V. (2026). Robotic reaming, multiple screw fixation results in significantly improved initial acetabular cup stability. HIP International. https://doi.org/10.1177/11207000251374145

Vancouver

Forlenza EM, Burnett RA, Espinoza Orías AA, Karas V. Robotic reaming, multiple screw fixation results in significantly improved initial acetabular cup stability. HIP International. 2026. doi:10.1177/11207000251374145

BibTeX
@article{Forlenza2026,
  author = {Forlenza, Enrico M and Burnett, Robert A and Espinoza Orías, Alejandro A and Karas, Vasili},
  title = {Robotic reaming, multiple screw fixation results in significantly improved initial acetabular cup stability},
  journal = {HIP International},
  year = {2026},
  doi = {10.1177/11207000251374145}
}
Kostenlos starten →