Long-term follow-up of total hip arthroplasty using polyethylene-ceramic composite (sandwich) liner
HIP International · 2024-07 · 34 (4) · 467-475
DOI: 10.1177/11207000241239624 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift HIP International, online erschienen am 26.03.2024. Verfasst von Daniel Rodríguez-Pérez und 5 weiteren Autor:innen. Beteiligte Einrichtung: Orthopaedic Surgery and Traumatology Department, Bellvitge University Hospital, L'Hospitalet de Llobregat, Barcelona, Spain. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip, Hip Prosthesis, Prosthesis Design, Ceramics. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 45 Literaturangaben.
Abstract (Originalsprache)
Introduction
Ceramic-on-ceramic bearings have been widely used since their introduction in the 1970s. First-generation ceramics have very high fracture (breakage) rates. To overcome this, in the 1990s, modular cotyloid insert designs were developed, consisting of a ceramic and polyethylene composite (sandwich-type) liner; however, high implant fracture rates were observed in the medium term. We aimed to estimate the cumulative incidences of revision surgery (implant failure) and implant failure due to fractures, survival rates (time-to-revision surgery and time-to-fracture), and the long-term clinical and radiological outcomes in our series.
Methods
This was an observational, longitudinal, ambispective, single-centre study based on patients who underwent primary total hip arthroplasty (THA) using a sandwich-type liner (Cerasul), only available in our tertiary hospital between January 1999 and December 2002. Cumulative incidences were estimated and the 95% confidence interval (95% CI) was calculated. The Kaplan-Meier method was used to assess the time-to-revision surgery and time-to-fracture.
Results
54 patients (49 men) were included, accounting for 59 sandwich-type linear implants. The mean (range) age was 47.4 (22–57) years. The primary THA indications were osteoarthritis (28 patients), osteonecrosis (14), childhood pathology sequelae (11), and inflammatory arthritis (6). The cumulative incidence of revision surgery by implants was 8.5% (5/59, 95% CI, 3.5–19.2%), 9.3% by patients (5/54, 95% CI, 4.0–19.9%), and 5.1% by implant fractures (3/59, 95%CI, 1.7–13.9%). The median (Interquartile Range, IQR) time-to-revision surgery was 158 (72.5–161) months, and the time to fracture was 182 (138–215) months. All primary THAs had good clinical and long-term survival outcomes. All implants had signs of solid fixation.
Conclusions
After a 20-year follow-up period, the polyethylene-ceramic sandwich-type liner showed a long survival rate and low cumulative incidence of implant fracture; however, implant fractures remain the main complication. Orthopaedic surgeons should be aware that some patients still have this type of prosthesis and must be capable of responding quickly if a fracture occurs.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Orthopaedic Surgery and Traumatology Department, Bellvitge University Hospital, L'Hospitalet de Llobregat, Barcelona, Spain.
- Clinical Research Support Unit (HUB-IDIBELL: Bellvitge University Hospital & Bellvitge Biomedical Research Institute), Clinical Pharmacology Department, Bellvitge University Hospital, L'Hospitalet de Llobregat, Barcelon…
- Pharmacology Unit, Department of Pathology and Experimental Therapeutics, School of Medicine and Health Sciences, IDIBELL, University of Barcelona, L'Hospitalet de Llobregat, Barcelona, Spain.
ORCID-Profile: Thiago Carnaval, Marcos-Del-Carmen Rodríguez
Zitieren
APA
Rodríguez-Pérez, D., Carnaval, T., Rodríguez, M. D., Coscujuela-Maña, A., Agulló, J. L., & Videla, S. (2024). Long-term follow-up of total hip arthroplasty using polyethylene-ceramic composite (sandwich) liner. HIP International, 34(4), 467-475. https://doi.org/10.1177/11207000241239624
Vancouver
Rodríguez-Pérez D, Carnaval T, Rodríguez MD, Coscujuela-Maña A, Agulló JL, Videla S. Long-term follow-up of total hip arthroplasty using polyethylene-ceramic composite (sandwich) liner. HIP International. 2024;34(4):467-475. doi:10.1177/11207000241239624
BibTeX
@article{RodrguezPrez2024,
author = {Rodríguez-Pérez, Daniel and Carnaval, Thiago and Rodríguez, Marcos-Del-Carmen and Coscujuela-Maña, Antonio and Agulló, José-Luis and Videla, Sebastián},
title = {Long-term follow-up of total hip arthroplasty using polyethylene-ceramic composite (sandwich) liner},
journal = {HIP International},
year = {2024},
volume = {34},
number = {4},
pages = {467--475},
doi = {10.1177/11207000241239624}
}
EUVIMED