Safety and efficacy of MicroPulse transscleral laser therapy with the revised P3 delivery device: A randomized controlled trial
European Journal of Ophthalmology · 2025
DOI: 10.1177/11206721251340102 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift European Journal of Ophthalmology, 2025. Verfasst von P Connor Lentz und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Mayo Clinic, Jacksonville, FL, USA. Hauptthemen laut MEDLINE: Sclera, Laser Coagulation, Glaucoma, Glaucoma, Open-Angle. Die Arbeit nennt 34 Literaturangaben.
Abstract (Originalsprache)
Objective
To evaluate the safety and efficacy of MicroPulse Transscleral Laser Therapy (MicroPulse TLT) with the Revised P3 ® Delivery Device.
Design
Prospective, single-surgeon, interventional study in a tertiary hospital setting.
Participants
Sixty-two eyes of 40 patients with refractory glaucoma.
Methods
All subjects received 360-degree MicroPulse TLT treatment (2500 mW, 31.33% duty cycle) over a total of 240 s (8 sweeps; n = 20), 300 s (10 sweeps; n = 22), or 200 s (10 sweeps; n = 20). Participants were followed up for 6 months to assess changes in intraocular pressure (IOP), glaucoma medication burden, and best-corrected visual acuity (BCVA). Surgical success was defined as a reduction of IOP ≥ 30% from baseline and between 6–18 mmHg at last follow-up with no additional anti-glaucoma medications or secondary reinterventions.
Results
Surgical success was achieved in 72.7% (300 s group), 75.0% (240 s group), and 80.0% (200 s group) of eyes. Mean IOP was significantly reduced ( p < 0.001) by 49.0% (300 s group), 45.66% (240 s group), and 39.8% (200 s group) at 6 months. With all groups combined, the mean number of glaucoma medications was significantly reduced by 17.8% ( p = 0.005), and no significant changes in BCVA were observed at 6 months. Complications were minimal, including four eyes with IOP spikes, one eye with hyphema, and one eye with cystoid macular edema (CME), all of which resolved. Two eyes of the total sample (3.2%) required secondary glaucoma surgery before 6 months.
Conclusions
Second-generation MicroPulse TLT at fluence values of 154–229 J/cm 2 appears to be a safe and effective option for IOP reduction in refractory glaucoma.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Ophthalmology, Mayo Clinic, Jacksonville, FL, USA.
- Department of Medicine, Medical College of Wisconsin, Milwaukee, WI, USA.
- Department of Ophthalmology, Tan Tock Seng Hospital, National Healthcare Group Eye Institute, Singapore, Singapore.
- Department of Ophthalmology, Woodlands Health, National Healthcare Group Eye Institute, Singapore, Singapore.
- Department of Ophthalmology, Casey Eye Institute, Portland, OR, USA.
ORCID-Profile: P Connor Lentz, Isabella V Wagner, Devesh Kumar, Abhimanyu Ahuja, Leticia Checo, Darby D Miller, Syril Dorairaj
Zitieren
APA
Lentz, P. C., Wagner, I. V., Kumar, D., Boopathiraj, N., Ang, B. C. H., Ahuja, A., Checo, L., Miller, D. D., & Dorairaj, S. (2025). Safety and efficacy of MicroPulse transscleral laser therapy with the revised P3 delivery device: A randomized controlled trial. European Journal of Ophthalmology. https://doi.org/10.1177/11206721251340102
Vancouver
Lentz PC, Wagner IV, Kumar D, Boopathiraj N, Ang BCH, Ahuja A, et al. Safety and efficacy of MicroPulse transscleral laser therapy with the revised P3 delivery device: A randomized controlled trial. European Journal of Ophthalmology. 2025. doi:10.1177/11206721251340102
BibTeX
@article{Lentz2025,
author = {Lentz, P Connor and Wagner, Isabella V and Kumar, Devesh and Boopathiraj, Nithya and Ang, Bryan C H and Ahuja, Abhimanyu and Checo, Leticia and Miller, Darby D and Dorairaj, Syril},
title = {Safety and efficacy of MicroPulse transscleral laser therapy with the revised P3 delivery device: A randomized controlled trial},
journal = {European Journal of Ophthalmology},
year = {2025},
doi = {10.1177/11206721251340102}
}
EUVIMED