Vergleichsstudie in der Zeitschrift European Journal of Ophthalmology, 2024. Verfasst von Jesús Merayo-Lloves und 3 weiteren Autor:innen. Beteiligte Einrichtung: Instituto Universitario Fernández Vega, Oviedo, Spain. Hauptthemen laut MEDLINE: Cornea, Dry Eye Syndromes. Die Arbeit wurde bisher 21-mal zitiert (Crossref) und nennt 26 Literaturangaben. Gefördert von: Brill Engines S.L,. Barcelona, Spain..
Abstract (Originalsprache)
Background
Corneal sensitivity can decrease by several ocular conditions, such as dry eye or refractive surgery, which favor ocular epithelial lesions and is measured using an esthesiometer. The study's primary objective was to demonstrate the efficacy and safety of the non-contact esthesiometer BRILL, which delivers air pulses to the corneal surface to assess corneal sensitivity.
Methods
A single-center, prospective, controlled pilot study was carried out in adult patients with healthy eyes and or with pathology. Corneal sensitivity measurements were made in triplicate for both eyes at three consecutive visits. The esthesiometer BRILL was used in all visits, and on the last visit, the contact esthesiometer Cochet-Bonnet was also used. The results of both devices were compared by transforming them into force values.
Results
54 subjects with a mean age of 50.43 (SD 16.55, interval 18–87), 77.78% women, were included. Comparing the forces applied by both esthesiometers in the healthy eyes, in the eyes with pathology in all the groups, and in the dry eyes showed significant differences, p = 0.03603, p = 0.00614, and p = 0.0001, respectively.
Conclusion
The BRILL esthesiometer proved to be an effective and safe tool for non-contact assessment of corneal sensitivity with operator-independent repeatability. The measurements had a good agreement and comparable range with the Cochet-Bonet aesthesiometer measurements in healthy and dry eyes but with no interchangeable values. This portable device can help ophthalmologists and optometrists to diagnose eye pathologies that cause decreased corneal sensitivity and to assess the efficacy of therapy and disease progression.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Instituto Universitario Fernández Vega, Oviedo, Spain.
- Universidad de Oviedo, Oviedo, Spain.
- Brill Engines, S.L., Barcelona, Spain.
ORCID-Profile: Carmen Gómez Martín, Javier Lozano-Sanroma
Förderung
- Brill Engines S.L,. Barcelona, Spain.
Zitieren
APA
Merayo-Lloves, J., Gómez Martín, C., Lozano-Sanroma, J., & Renedo Laguna, C. (2024). Assessment and safety of the new esthesiometer BRILL: Comparison with the Cochet-Bonnet Esthesiometer. European Journal of Ophthalmology. https://doi.org/10.1177/11206721231210754
Vancouver
Merayo-Lloves J, Gómez Martín C, Lozano-Sanroma J, Renedo Laguna C. Assessment and safety of the new esthesiometer BRILL: Comparison with the Cochet-Bonnet Esthesiometer. European Journal of Ophthalmology. 2024. doi:10.1177/11206721231210754
BibTeX
@article{MerayoLloves2024,
author = {Merayo-Lloves, Jesús and Gómez Martín, Carmen and Lozano-Sanroma, Javier and Renedo Laguna, Concepción},
title = {Assessment and safety of the new esthesiometer BRILL: Comparison with the Cochet-Bonnet Esthesiometer},
journal = {European Journal of Ophthalmology},
year = {2024},
doi = {10.1177/11206721231210754}
}
EUVIMED