Evaluationsstudie in der Zeitschrift Surgical Infections, online erschienen am 01.04.2026. Verfasst von Katherine Dziak. Beteiligte Einrichtung: Research & Development, Mölnlycke Health Care, Peachtree Corners, Georgia, USA. Hauptthemen laut MEDLINE: Gloves, Surgical, Materials Testing. Die Arbeit nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background
Surgical glove breaches may compromise the aseptic barrier and pose risks to surgical staff. Double gloving with an indicator system improves breach detection compared with similar colored double gloves or single gloving. This study evaluates the performance of market-leading indicator gloves for breach detection.
Methods
Liquid was introduced between layers of breach indicating polyisoprene double gloves (Brands A [six styles], B [three styles], and C [two styles]). Indication area was measured using a calibrated camera at two time points (immediate = 9.8 sec; final = 49 sec). Performance differences between styles and brands were analyzed using Games–Howell and Mood’s tests (95% confidence interval).
Results
Brand A had a larger and faster immediate median indication area (544 mm 2 ; 55.5 mm 2 /sec; p < 0.05), compared with Brand B (61 mm 2 ; 6.2 mm 2 /sec) and Brand C (115 mm 2 ; 11.7 mm 2 /sec); this was 9.0 and 4.7 times faster than Brands B and C, respectively. Brand A also displayed a larger final median indication area (597 mm 2 ; p < 0.05), compared with Brands B (124 mm 2 ) and C (314 mm 2 ), with a 4.8 and 1.9 times larger area, respectively. Brand A immediately reached 91% of its final indication area compared with only 49% and 37% for Brands B and C, respectively.
Conclusions
Indicator glove performance influences breach visualization. Brand A gloves, which have unique surface characteristics, displayed the fastest and largest breach indication among the brands tested. Faster and larger breach indication enables rapid detection and response, protecting the aseptic barrier between the wearer and patient.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Research & Development, Mölnlycke Health Care, Peachtree Corners, Georgia, USA.
ORCID-Profile: Katherine Dziak
Zitieren
APA
Dziak, K. (2026). Detecting Surgical Glove Breaches: Evaluation of Double Glove Indicator System Performance Using an In Vitro Glove Perforation Method. Surgical Infections, 27(7), 483-488. https://doi.org/10.1177/10962964261430395
Vancouver
Dziak K. Detecting Surgical Glove Breaches: Evaluation of Double Glove Indicator System Performance Using an In Vitro Glove Perforation Method. Surgical Infections. 2026;27(7):483-488. doi:10.1177/10962964261430395
BibTeX
@article{Dziak2026,
author = {Dziak, Katherine},
title = {Detecting Surgical Glove Breaches: Evaluation of Double Glove Indicator System Performance Using an In Vitro Glove Perforation Method},
journal = {Surgical Infections},
year = {2026},
volume = {27},
number = {7},
pages = {483--488},
doi = {10.1177/10962964261430395}
}
EUVIMED