Risk assessment of pneumatic tube systems for in-hospital transportation of biotherapeutics
Journal of Oncology Pharmacy Practice · 2025
DOI: 10.1177/10781552251378729 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Oncology Pharmacy Practice, 2025. Verfasst von Michaela Cohrs und 7 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory of General Biochemistry and Physical Pharmacy, Ghent University, Ghent, Belgium. Hauptthemen laut MEDLINE: Transportation, Biological Products. Die Arbeit nennt 32 Literaturangaben. Gefördert von: Research Foundation Flanders - FWO.
Abstract (Originalsprache)
Introduction
Use of Pneumatic tube systems (PTS's) poses a risk to inherently unstable protein therapeutics. Mechanical stress caused by those systems can lead to aggregation, compromising product safety and quality. However, PTS's are widely used in hospitals, and contradictory effects on protein stability have been reported.
Methods
We studied whether mechanical stress levels differ between hospitals and routes, potentially leading to different findings regarding protein aggregation. By transporting a mechanical sensor on nine different PTS routes in seven different hospitals, we quantified the accelerations that protein therapeutics experience in a specific PTS setup.
Results
It was found that the maximum acceleration within the different PTS setups was similar. However, transportation times differed up to 4-fold, leading to varying amounts of cumulative stress.
Conclusion
The cumulative and maximum stress quantified can be compared to stress found in studies investigating protein stability, which justifies the usage of mechanical sensors for risk evaluation. The differences in stress found underscore the importance of evaluating handling practices individually in each hospital, for each therapeutic protein.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Laboratory of General Biochemistry and Physical Pharmacy, Ghent University, Ghent, Belgium.
- Department of Hospital Pharmacy, Meander Medical Centre, Amersfoort, Netherlands.
- Pharmacy Department, Ghent University Hospital, Ghent, Belgium.
- Biopharmaceutical Technology, TUM School of Life Sciences, Technical University of Munich, Freising, Germany.
ORCID-Profile: Michaela Cohrs, Kevin Braeckmans, Stefaan De Smedt, Hristo L Svilenov, Tiene Bauters
Förderung
- Research Foundation Flanders - FWO
Zitieren
APA
Cohrs, M., Borgonje, P. E., Clottens, N., Ramaut, P., Braeckmans, K., De Smedt, S., Svilenov, H. L., & Bauters, T. (2025). Risk assessment of pneumatic tube systems for in-hospital transportation of biotherapeutics. Journal of Oncology Pharmacy Practice. https://doi.org/10.1177/10781552251378729
Vancouver
Cohrs M, Borgonje PE, Clottens N, Ramaut P, Braeckmans K, De Smedt S, et al. Risk assessment of pneumatic tube systems for in-hospital transportation of biotherapeutics. Journal of Oncology Pharmacy Practice. 2025. doi:10.1177/10781552251378729
BibTeX
@article{Cohrs2025,
author = {Cohrs, Michaela and Borgonje, Paula E and Clottens, Nele and Ramaut, Pieter and Braeckmans, Kevin and De Smedt, Stefaan and Svilenov, Hristo L and Bauters, Tiene},
title = {Risk assessment of pneumatic tube systems for in-hospital transportation of biotherapeutics},
journal = {Journal of Oncology Pharmacy Practice},
year = {2025},
doi = {10.1177/10781552251378729}
}
EUVIMED