Accuracy of a Patient-Specific Total Ankle Arthroplasty Instrumentation
Foot & Ankle International · 2023-11 · 44 (11) · 1150-1157
DOI: 10.1177/10711007231194049 ↗Kurz gefasst
Kommentar in der Zeitschrift Foot & Ankle International, online erschienen am 20.09.2023. Verfasst von Pierre-Marc April und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedics, Civic Hospital, Ottawa, Canada. Hauptthemen laut MEDLINE: Ankle, Arthroplasty, Replacement, Ankle. Die Arbeit wurde bisher 16-mal zitiert (Crossref) und nennt 41 Literaturangaben.
Abstract (Originalsprache)
Background
Total ankle arthroplasty (TAA) is a preferred surgical option for end-stage ankle osteoarthritis; however, it is a demanding procedure with a higher historical rate of revision compared with ankle fusion. Patient-specific instrumentation (PSI) has been introduced to optimize prosthesis alignment and theoretically overall improve TAA outcomes. The goal of this study is to report on the experience and surgical outcomes of one implant with specific evaluation of the accuracy and reproducibility of the system with respect to prosthesis alignment and prediction of implant size.
Methods
A retrospective, multicentered study involving 4 foot and ankle fellowship–trained orthopaedic surgeon’s patients undergoing TAA between January 1, 2015, and December 31, 2018, using the PROPHECY PSI system.
Results
80 TAA procedures were performed. On average the postoperative tibial component alignment was 89.9 (range, 86.1-96.5) degrees in the coronal plane, with a mean sagittal alignment of 88.1 (range, 81.3-96.7) degrees. The mean deviation from neutral sagittal alignment improved from 4.9 ± 3.9 degrees preoperatively to 2.7 ± 1.7 degrees postoperatively, whereas the mean coronal alignment improved from 3.3 ± 2.5 degrees to 1.3 ± 1.1 degrees. The PSI software correctly determined the tibial implant size in 70 patients (89%). Prediction of talar implant sizing was less accurate than the tibial component, with 56 patients (71%) using the predicted sized implant. The overall implant survival at a mean follow-up of 45 months (range, 27-76) was 97.5%.
Conclusion
We found that this PSI system accurately and reliably assisted in implant total ankle prosthesis positioning within a clinically acceptable margin and without significant outliers. Prediction of implant size was not as accurate as component orientation. Level of Evidence: Level III, retrospective study.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopaedics, Civic Hospital, Ottawa, Canada.
- Department of Orthopaedics, CHUS, Sherbrooke, Canada.
ORCID-Profile: Pierre-Marc April
Zitieren
APA
April, P. M., Locke, E., Champagne, P. H., Angers, M., Martinez-Gomez, A. P., Seidel, A., Jibri, Z., Lalonde, K. A., & Meulenkamp, B. (2023). Accuracy of a Patient-Specific Total Ankle Arthroplasty Instrumentation. Foot & Ankle International, 44(11), 1150-1157. https://doi.org/10.1177/10711007231194049
Vancouver
April PM, Locke E, Champagne PH, Angers M, Martinez-Gomez AP, Seidel A, et al. Accuracy of a Patient-Specific Total Ankle Arthroplasty Instrumentation. Foot & Ankle International. 2023;44(11):1150-1157. doi:10.1177/10711007231194049
BibTeX
@article{April2023,
author = {April, Pierre-Marc and Locke, Eric and Champagne, Philippe-Hugo and Angers, Magalie and Martinez-Gomez, Adriana Patricia and Seidel, Angela and Jibri, Zaid and Lalonde, Karl-Andre and Meulenkamp, Brad},
title = {Accuracy of a Patient-Specific Total Ankle Arthroplasty Instrumentation},
journal = {Foot & Ankle International},
year = {2023},
volume = {44},
number = {11},
pages = {1150--1157},
doi = {10.1177/10711007231194049}
}
EUVIMED