Raising the awareness for insufficient oxygen delivery from self-inflating resuscitation bags lacking expiratory valve during preoxygenation

Yu-Yeung Yip, Christopher Pak-To Lee, Jonathan Chun-Hei Cheung

Hong Kong Journal of Emergency Medicine · 2023-01 · 30 (1) · 61-63

DOI: 10.1177/10249079221096918 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Hong Kong Journal of Emergency Medicine, online erschienen am 13.05.2022. Verfasst von Yu-Yeung Yip und 2 weiteren Autor:innen. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 2 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

We recently read an interesting study which demonstrated that self-inflating resuscitation bag (SIRB) lacking expiratory valve has unreliable performance in oxygen delivery during spontaneous breathing mimicked by mechanical lung simulator. It was postulated that the absence of an expiratory valve and the resulting air entrainment via the exhaust port accounts for the poor oxygen delivery performance. The current disposable SIRB in-use in our institutions (Med-Rescuer Disposable BVM Resuscitator 4000, BLS Systems Limited, ON, Canada) has a duckbill valve but no expiratory valve. Safety concerns regarding its oxygen delivery performance during spontaneous breathing were raised, as this SIRB was commonly used to preoxygenate critically ill patient with potentially transmissible respiratory infection (e.g. COVID-19) before tracheal intubation. We therefore performed an experiment on this SIRB using one of us as a healthy volunteer. Our small experiment demonstrated that air entrainment could occur via the exhaust port and affect oxygen delivery performance. Our experiment also demonstrated that attaching a positive end-expiratory pressure (PEEP) valve to the exhaust port improves the oxygen delivery performance. The findings of this experiment were sent to the relevant department of our institutions for safety consideration.

Zitieren

APA

Yip, Y., Lee, C. P., & Cheung, J. C. (2023). Raising the awareness for insufficient oxygen delivery from self-inflating resuscitation bags lacking expiratory valve during preoxygenation. Hong Kong Journal of Emergency Medicine, 30(1), 61-63. https://doi.org/10.1177/10249079221096918

Vancouver

Yip Y, Lee CP, Cheung JC. Raising the awareness for insufficient oxygen delivery from self-inflating resuscitation bags lacking expiratory valve during preoxygenation. Hong Kong Journal of Emergency Medicine. 2023;30(1):61-63. doi:10.1177/10249079221096918

BibTeX
@article{Yip2023,
  author = {Yip, Yu-Yeung and Lee, Christopher Pak-To and Cheung, Jonathan Chun-Hei},
  title = {Raising the awareness for insufficient oxygen delivery from self-inflating resuscitation bags lacking expiratory valve during preoxygenation},
  journal = {Hong Kong Journal of Emergency Medicine},
  year = {2023},
  volume = {30},
  number = {1},
  pages = {61--63},
  doi = {10.1177/10249079221096918}
}