Reusing high-value consumables in shoulder arthroscopy: feasibility and overlooked complications

Lu Yang, Li Xiang, Li Huafeng, Jin Rilong, Hu Yihe iD

Journal of Orthopaedic Surgery · 2025

DOI: 10.1177/10225536251387309 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Orthopaedic Surgery, 2025. Verfasst von Lu Yang und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedics, First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou, China. Hauptthemen laut MEDLINE: Arthroscopy, Equipment Reuse, Shoulder Joint, Disposable Equipment. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

In many countries, disposable high-value consumables are sterilized and reused for economic reasons. We aimed to study the feasibility and ways to reduce potential risks of reusing these consumables in shoulder arthroscopy.

Methods

We reported cases with complications potentially related to reusing high-value consumables. We evaluated the outflow water temperature from new and reused radiofrequency (RF) devices of different brands and under varying surgical practices. We also conducted fatigue tests on suture-passing needles from different brands subjected to various disinfection methods.

Results

Reused RF devices from MECHAN demonstrated significantly higher peak temperature (PT) (68.3 ± 2.3°C vs 63.9 ± 3.0°C, P < 0.001) and peak sustaining temperature (PST) (62.6 ± 2.3°C vs 58.5 ± 3.1°C, P < 0.001) compared to new devices. The mean flow rate of MECHAN devices was higher than that of DePuy (63.2 ± 8.2 ml/min vs 42.0 ± 11.1 ml/min, P < 0.001). Ethylene oxide disinfection did not significantly affect the excitation times before fatigue fracture of Arthrex (291.6 ± 76.9 vs 305.1 ± 86.6, P = 0.717) and DePuy (173.3 ± 26.8 vs 174.6 ± 31.7, P = 0.922) suture-passing needles. High-pressure steam disinfection also did not markedly affect the fatigue time of Arthrex (291.6 ± 76.9 vs 292.0 ± 65.7, P = 0.990) and DePuy (173.3 ± 26.8 vs 168.8 ± 36.1, P = 0.755) suture-passing needles.

Conclusions

For both RF devices and suture-passing needles, being reused after ethylene oxide disinfection for appropriate times can be feasible and safe if surgeons and disinfection suppliers exercise caution.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ArthroscopyEquipment ReuseShoulder JointDisposable EquipmentHumansFeasibility StudiesDisinfection

Stichwörter der Autor:innen

complicationsdermal burnsreusing consumablesshoulder arthroscopysuture-passing needle breaking

Autor:innen und Institutionen

  • Department of Orthopedics, First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou, China.
  • Department of Engineering Mechanics, Zhejiang University, Hangzhou, China.

ORCID-Profile: Hu Yihe

Zitieren

APA

Yang, L., Xiang, L., Huafeng, L., Rilong, J., & Yihe, H. (2025). Reusing high-value consumables in shoulder arthroscopy: feasibility and overlooked complications. Journal of Orthopaedic Surgery. https://doi.org/10.1177/10225536251387309

Vancouver

Yang L, Xiang L, Huafeng L, Rilong J, Yihe H. Reusing high-value consumables in shoulder arthroscopy: feasibility and overlooked complications. Journal of Orthopaedic Surgery. 2025. doi:10.1177/10225536251387309

BibTeX
@article{Yang2025,
  author = {Yang, Lu and Xiang, Li and Huafeng, Li and Rilong, Jin and Yihe, Hu},
  title = {Reusing high-value consumables in shoulder arthroscopy: feasibility and overlooked complications},
  journal = {Journal of Orthopaedic Surgery},
  year = {2025},
  doi = {10.1177/10225536251387309}
}
Kostenlos starten →