Reusing high-value consumables in shoulder arthroscopy: feasibility and overlooked complications
Journal of Orthopaedic Surgery · 2025
DOI: 10.1177/10225536251387309 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Orthopaedic Surgery, 2025. Verfasst von Lu Yang und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedics, First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou, China. Hauptthemen laut MEDLINE: Arthroscopy, Equipment Reuse, Shoulder Joint, Disposable Equipment. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background
In many countries, disposable high-value consumables are sterilized and reused for economic reasons. We aimed to study the feasibility and ways to reduce potential risks of reusing these consumables in shoulder arthroscopy.
Methods
We reported cases with complications potentially related to reusing high-value consumables. We evaluated the outflow water temperature from new and reused radiofrequency (RF) devices of different brands and under varying surgical practices. We also conducted fatigue tests on suture-passing needles from different brands subjected to various disinfection methods.
Results
Reused RF devices from MECHAN demonstrated significantly higher peak temperature (PT) (68.3 ± 2.3°C vs 63.9 ± 3.0°C, P < 0.001) and peak sustaining temperature (PST) (62.6 ± 2.3°C vs 58.5 ± 3.1°C, P < 0.001) compared to new devices. The mean flow rate of MECHAN devices was higher than that of DePuy (63.2 ± 8.2 ml/min vs 42.0 ± 11.1 ml/min, P < 0.001). Ethylene oxide disinfection did not significantly affect the excitation times before fatigue fracture of Arthrex (291.6 ± 76.9 vs 305.1 ± 86.6, P = 0.717) and DePuy (173.3 ± 26.8 vs 174.6 ± 31.7, P = 0.922) suture-passing needles. High-pressure steam disinfection also did not markedly affect the fatigue time of Arthrex (291.6 ± 76.9 vs 292.0 ± 65.7, P = 0.990) and DePuy (173.3 ± 26.8 vs 168.8 ± 36.1, P = 0.755) suture-passing needles.
Conclusions
For both RF devices and suture-passing needles, being reused after ethylene oxide disinfection for appropriate times can be feasible and safe if surgeons and disinfection suppliers exercise caution.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopedics, First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou, China.
- Department of Engineering Mechanics, Zhejiang University, Hangzhou, China.
ORCID-Profile: Hu Yihe
Zitieren
APA
Yang, L., Xiang, L., Huafeng, L., Rilong, J., & Yihe, H. (2025). Reusing high-value consumables in shoulder arthroscopy: feasibility and overlooked complications. Journal of Orthopaedic Surgery. https://doi.org/10.1177/10225536251387309
Vancouver
Yang L, Xiang L, Huafeng L, Rilong J, Yihe H. Reusing high-value consumables in shoulder arthroscopy: feasibility and overlooked complications. Journal of Orthopaedic Surgery. 2025. doi:10.1177/10225536251387309
BibTeX
@article{Yang2025,
author = {Yang, Lu and Xiang, Li and Huafeng, Li and Rilong, Jin and Yihe, Hu},
title = {Reusing high-value consumables in shoulder arthroscopy: feasibility and overlooked complications},
journal = {Journal of Orthopaedic Surgery},
year = {2025},
doi = {10.1177/10225536251387309}
}
EUVIMED